FCU menus
Setting the averaging interval
Entering an averaging interval enables the data quantity which accumulates
during a measurement to be reduced.
An averaging interval of > 0 min causes only the mean value of all
measurements completed within this interval to be saved and printed out.
The 0 min setting causes the averaging function to be deactivated. Each
measured value is immediately saved and printed out.
The values 0-1440 minutes are possible for the averaging interval.
The current, non-averaged measured value is always shown in the FCU
display.
memory
MEMORY 0.0 %
Averaging interval
ok
start
Averaging interval:
x
Averaging interval:
x
or
Averaging interval:
x [min]
ok
start
Confirm entries.
stop
Abort and exit without saving.