background image

6

Thank you very much for selecting our products

!

For your better use of this instrument, you are kindly requested to

read this manual carefully before you use it, so as to understand its
major performances and operation method.

Caution

:

It is not allowed to remove the wiring and directly switch off the

power supply during the measurement.

For the no-load voltage regulating transformer, it is not allowed to

switch the tapping switch during the measurement.

In case of a sudden power failure during the measurement, this

instrument can discharge automatically. Please do not remove
the wiring immediately. Wait for at least 30 seconds before the
connection can be removed.

The three-phase measurement applies to the windings of YN
connection. For the windings of yn connection, since difference
exists between the three-phase and single-phase measurement
results due to the effect of the resistance of the copper bus
connected, the single-phase measurement is recommended.

Caution: The three-phase measurement result doses not include

the resistance value of the neutral lead. Therefore, it is
suggested to at least verify one data using the single-phase
testing method for each phase after completing the
measurement of three-phase resistance value, so as to
determine if the neutral lead is normal.

I. General

This fast DC resistance tester (hereinafter referred to as DC resistance

tester) is the product of the latest generation for measuring the DC resistance
of transformer and is optimally designed for measuring the DC resistance of
the three-phase winding of the large-capacity transformer. It can measure the
DC resistance of the three-phase winding at the same time. In case of the
on-load voltage regulating transformer, it is not required to discharge but to
directly adjust the tapping switch. The measurement time is one third of the
traditional single-phase measurement, greatly shortening the working time and
labor intensity. As the DC resistor tester is provided with large-screen LCD and
the graphical interfaces in Chinese, the operation is visual, clear and easy.

Содержание HTZZ-S10A

Страница 1: ...manual You can find the changes in the appendix Sorry for the inconvenience If you have further questions welcome to contact with our service department The input output terminals and the test column...

Страница 2: ...and prevent the product or other relevant subassembly to damage In order to avoid possible danger this product can only be used within the prescribed scope Only qualified technician can carry out main...

Страница 3: ...d the instruction manual to acquire information about the ratings Do Not Operate without Covers Do not operate this product when covers or panels removed Use Proper Fuse Only use the fuse with type an...

Страница 4: ...Security Terms Warning indicates that death or severe personal injury may result if proper precautions are not taken Caution indicates that property damage may result if proper precautions are not tak...

Страница 5: ...I General 6 II Performance Index 7 III Panel Description 7 IV Testing and Operating Method 9 V Precautions 14 VI Troubleshooting 15 VII Transportation and Storage 15 VIII Unpacking and Inspection 16 I...

Страница 6: ...hase measurement is recommended Caution The three phase measurement result doses not include the resistance value of the neutral lead Therefore it is suggested to at least verify one data using the si...

Страница 7: ...Single phase 1A 5A 10A 20A Three phase 1A 5A 10A Measurement scope Single phase 1A 10 m 20 5A 1 m 4 10A 1 m 2 20A 1 m 1 Three phase 1A 10 m 6 each phase 5A 1 m 1 each phase 10A 1 m 0 6 each phase Mea...

Страница 8: ...the first channel Uc and Uo are the input of the second channel 3 Panel printer Prints the output measurement results 4 LCD displays the measurement results and operation prompting information with t...

Страница 9: ...easure 1 4 0 2 5 1 Set key press Set the current date and time View key press View the data stored including measurement results testing date tapping number and etc Mode key press Sets the measurement...

Страница 10: ...nters into the interface as follows 2 0 0 7 0 1 0 1 Tap R 0 0 A Store The battery is charging please wait a moment Current tap 01 Print Re test Exit 1 4 0 2 5 1 At this time the current of the winding...

Страница 11: ...tton to modify it Then press Store and Print button to save or print the measurement result In case of any doubt about the data press Retest button to carry out the measurement and calculation again W...

Страница 12: ...the tester to enter into its measuring state and to start to supply power to the winding and then it enters into the interface as follows 2 0 0 7 0 1 0 1 Tap A 0 0 A B 0 0 A Store C 0 0 A The battery...

Страница 13: ...rement result After the Tap button is pressed down the multi functional key on the right becomes OK and Cancel and the tap setting dialogue box is ejected After setting press the OK button to save it...

Страница 14: ...g may be removed V Precautions 1 When measuring the no load voltage changing transformer be sure to exit from the discharge before switching the tapping switch Do not switch the tapping switch before...

Страница 15: ...t VII Transportation and Storage Transportation When transporting the equipment it is suggested using the instrument packing wood boxes and damping materials of our company to avoid any unnecessary da...

Страница 16: ...y shortage please make contact with our company We will give serve you as soon as possible IX Others The warranty period for the whole set of this product is one year and Three guarantees repair retur...

Отзывы: