6
I. Overview
In power system and industrial and mining enterprises have a
lot of oil filled electrical equipment, transformer oil, and the acid
value of turbine oil is a regular inspection items shall be measured.
Insulation oleic acid value automatic tester used in transformer oil,
turbine oleic acid determination instrument, adopt computer control,
automatic electrical, optical, mechanical and chemical integration
design, completely replaces manual operation, raised the working
efficiency, reduces the labor intensity of workers, minimize the
organic solvents and chemicals to the harm of human body. The
operation is simple, easy to use, strong anti-interference ability,
stable and reliable
.
The acid value is the number of milligrams of potassium
hydroxide in 1 gram of oil in the acidic substances, with mgKOH / g
expressed; that it is one of the indicators of the weight of oil should
be strictly controlled. This tester according to GB / T264-83
standard developed and produce for acid value measurement,
analysis of transformer oil, turbine oil and anti-fuel oil samples. This
product through a combination of mechanical, optical and
electronic technology, use advanced microprocessors, capable of
automatically switching multi-sample, titration to determine the
titration end point, print the results and other functions, the system