16
(2) blank test
: the interface is as follows
Used for measuring the acid value of the extract. If changed
the extract liquid, blank experiment must do first before sample test.
Method is as follows: Put one empty cup into test cup position,
place one stir, and then select the test cup, for example " 1 cup",
select " sample weight" corresponds to the test cup and with "▲",
"▼" button adjust it to "0.000", then the sample weight shows
"blank", as shown in the figure. After press "start" button, the
machine will automatically blank test operation. (Note: to observe
the neutralization liquid titration speed about every drop 1 second)
and then take out the cup of blank test.
(3)Cur.Hue:
is value of the color sensor, instrument automatic
detect and display.
(4)End Hue:
basis of titration end point, is refers to the color
sensor value at the end of test, that is to say, when the current