background image

4

voltage U is the product of input reference voltage Uref and K. As K does not affect the

measurement of angle or current magnitude, it can be set to 1, directly displayed as U1. Note that

the correct measurements cannot be obtained when there is no U1 input. For the stored data, the

voltage transformation ratio can be modified, but the modified value only influences the currently

displayed/printed data and cannot be saved.

Test phase difference: the phase difference of the leading fundamental voltage at the

fundamental current includes the compensation angle. MOA performance can be evaluated

according to the angle, and shall be evaluated by deducting the interference angle in case of

interphase interference.

Waveform display

Full voltage waveform

Full current waveform

Resistive current waveform

4. Click “Data View” to display the menu as shown in Figure 7:

0000-A January 5, 2015 08:08:08 001-001

Full V

0.000KV

Full I

0.000mA

Fund V

0.000KV

Res I

0.000mA

Cap I

0.000mA Res fund I

0.000mA

Freq

50.00Hz

Res I3

0.000mA

A Power

0.000W

Res I5

0.000mA

R Power

0.000W

Res I7

0.000mA

Ratio

0.000

Res IMax

0.000mA

Angle

0.000

Diff Phase

0.00

Prev

Next

Y

Print

N

Esc

Figure 7 Data View Main Menu

5. Printout

To print, directly operate according to the screen prompt. In order to facilitate the user to

analyze and save the test data, the instrument saves 100 sets of test data for the user to print.

(After measurement, the user can save the data as required.)

6. Instrument calibration

Enter the password to enter the instrument calibration. Enter the instrument calibration

only when the instrument accuracy is wrong. Do not enter at other times.

7.

Time calibration

January 5, 2015 08:08:08

↑ Change digits ↓ Move cursor

VI. Measurement Principle and Data Analysis

:

The instrument inputs PT secondary voltage as a reference signal and inputs MOA current

signal to obtain the fundamental voltage U1, the fundamental current peak and the current and

voltage angle Φ (Figure 8) by Fourier transformation. The voltage cophase component is resistive

current fundamental value (Ir1p), and the quadrature component is apacitive current fundamental

value (Ic1p):

Ir1p=Ix1pCOSΦ

Ic1p=Ix1pSINΦ

Содержание HTYB-3H

Страница 1: ...u can find the changes in the appendix Sorry for the inconvenience If you have further questions welcome to contact with our service department The input output terminals and the test column may bring...

Страница 2: ...this product can only be used within the prescribed scope Only qualified technician can carry out maintenance or repair work To avoid fire and personal injury Use Proper Power Cord Only use the power...

Страница 3: ...o not operate this product when covers or panels removed Use Proper Fuse Only use the fuse with type and rating specified for the product Avoid Touching Bare Circuit and Charged Metal Do not touch the...

Страница 4: ...ecurity Terms Warning indicates that death or severe personal injury may result if proper precautions are not taken Caution indicates that property damage may result if proper precautions are not take...

Страница 5: ...able of Contents I Overview 1 II Structure Diagram for Instrument Panel 1 III Main Technical Indexes 1 IV Wiring Diagrams 1 V Instrument Operation 2 VI Measurement Principle and Data Analysis 4 VII Pr...

Страница 6: ...ests II Structure Diagram for Instrument Panel Figure 1 III Main Technical Indexes Range of reference voltage input peak 10V 200V Customizable 0 250V Measurement range of full leakage current peak 100...

Страница 7: ...r and the ground Boosting value 10KV 1 5 35KV 3 2 Online wiring diagram live test Figure 3 In the online measurement the voltage signal input is connected to the secondary side of PT located at the sa...

Страница 8: ...inding For example for the testing transformer with AC output rated voltage of 50KV the rated voltage of the measuring instrument winding is usually 100V so the transformation ratio is 50KV 100V 500 D...

Страница 9: ...s fund I 0 000mA Freq 50 00Hz Res I3 0 000mA A Power 0 000W Res I5 0 000mA R Power 0 000W Res I7 0 000mA Ratio 0 000 Res IMax 0 000mA Angle 0 000 Diff Phase 0 00 Prev Next Y Print N Esc Figure 7 Data...

Страница 10: ...rference can be considered when MOA performance is evaluated Based on B phase the reduced value of A phase is basically equal to the increased value of C phase according to the symmetry of interphase...

Страница 11: ...t To judge whether the valve disc of the metal oxide arrester is aged or affected with damp usually observe the change of resistive leakage current flowing through the zinc oxide valve disc in the nor...

Страница 12: ...7 you understand the change to the resistive leakage current of the metal oxide arrester you can predict whether the above failures occur...

Отзывы: