background image

9

500V. Select the required voltage grade and the indicator light will

show the selected voltage grade. Press the high voltage “start/stop”

button and the indicator light for high voltage will glow. The stable

value displaying on the LCD multiplied by 10 is the measured

insulation resistance value. When the tested insulation resistance

value exceeds the upper limit range of the meter, the screen will

only display “1” on the left-most position. Pressing “start/stop”

button will turn off the high voltage. Pressing down the “ON/OFF”

button will turn off the power supply.

Notes: As measure, accompanied with absorption and

polarization occurring, Insulation reading gradually drifts to large

values or bounces up and down, it is normal.

2. Explanation for the wiring terminals

When measuring the insulation resistance you should connect

the “L” line and the measured object to the current-carrying part.

The grounding “E” should be connected to the shell of the

measured object or the earth. Shield “G” is connected to the

protection parts or the other parts which are not being measured of

the measured object to avoid the error caused by shell leakage.

When measuring the insulation resistance among components of

electronic products, the “L” terminal and the “E” terminal can be

connected to any line. When measuring the resistance between

phases of electrical generator, you can choose two among the

Содержание HT2672

Страница 1: ...You can find the changes in the appendix Sorry for the inconvenience If you have further questions welcome to contact with our service department The input output terminals and the test column may bring voltage when you plug draw the test wire or power outlet they will cause electric spark PLEASE CAUTION RISK OF ELECTRICAL SHOCK Company Address T4 No 41 High tech 2 Road East Lake High tech Develop...

Страница 2: ...er relevant subassembly to damage In order to avoid possible danger this product can only be used within the prescribed scope Only qualified technician can carry out maintenance or repair work To avoid fire and personal injury Use Proper Power Cord Only use the power wire supplied by the product or meet the specification of this produce Connect and Disconnect Correctly When the test wire is connec...

Страница 3: ...quire information about the ratings Do Not Operate without Covers Do not operate this product when covers or panels removed Use Proper Fuse Only use the fuse with type and rating specified for the product Avoid Touching Bare Circuit and Charged Metal Do not touch the bare connection points and parts of energized equipment Do Not Operate with Suspicious Failures If you encounter operating failure d...

Страница 4: ... Security Terms Warning indicates that death or severe personal injury may result if proper precautions are not taken Caution indicates that property damage may result if proper precautions are not taken ...

Страница 5: ...5 Contents I Introduction 6 II Function and Features 7 III Technique Specification 7 IV Operation Instructions 8 V Notes 10 ...

Страница 6: ... a current from pole E to pole L which converted by I V and computed by the divider The value of insulation resistance will display on the LCD finally 2 Block Diagram of Circuit As shown in the following diagram 3 Applications The meter is essential and common device in power industry post and telecommunications communications mechanical and electrical installation and maintenance and industrial s...

Страница 7: ...ctromagnetic field For the measurement of capacitive specimen since the short circuit current is greater than 1 6mA it s easy to rise the test voltage to the rating of output voltage For the measurement of low resistance it will not affect the test by using scaling method 2 The meter is battery powered and the measure range is converted automatically The easy to read panel and LCD display make mea...

Страница 8: ...m output voltage and short circuit current Voltage Load 5000V 40MΩ Voltage drop Approximately 10 Short circuit current 1 6mA 4 Power supply power dissipation DC 8 1 5V AA R6 battery AC 220V 50Hz Power dissipation quiescent power dissipation 160mW Maximum power 2 5W 5 Dimensions and weight Dimensions 235mm L 200mm W 135mm D Weight 1 4kg IV Operation Instructions 1 Measurement Procedure Turn the pow...

Страница 9: ...nied with absorption and polarization occurring Insulation reading gradually drifts to large values or bounces up and down it is normal 2 Explanation for the wiring terminals When measuring the insulation resistance you should connect the L line and the measured object to the current carrying part The grounding E should be connected to the shell of the measured object or the earth Shield G is conn...

Страница 10: ...hat may endanger safety of your body and the device The DC voltage between the E terminal and L terminal is high after turning on the high voltage Make sure any parts of your body should not touch the terminals during measuring 3 This meter uses AC and DC power supply When connect to AC power supply the meter will prior use AC power supply when not connect to AC power supply battery will power met...

Отзывы: