User Manual
HV Diagnostics
Page 28
Rev 9.0
Step
Procedure
(Set Manual Test Parameters)
Example
MS 6:
Sheath Fault
Location
Select one of the following Pulse/Periods:
•
1:3 / 4 s
•
1:5 / 4 s
•
1:5 / 6 s
•
1:9 / 6 s
For example: 1 second on and 3 seconds off,
every 4 seconds.
MS 7:
SETUP:
Duration
(Not applicable
Vacuum Bottle
Test)
To modify the duration, rotate navigation knob (5).
To accept value, push in knob.
•
Min. test duration = 1 minute
•
Max. test duration = 24 hours
To return to “Manual Mode” screen, select “OK”
MS 8:
Arc Management
Mode
Rotate navigation knob (5) until the field on bottom
of the screen is highlighted.
To change the mode, push in the knob. One of the
following modes will be displayed:
•
Trip out on Arc
•
Burn on Arc
MS 8:
Preset Test
Voltage:
(optional-voltage
can be set once
test has been
initiated!)
Entering the test voltage before activating the
manual mode test “START” is
optional
.
In manual mode, voltage can be set once test has
been initiated!
To set the test voltage before activating the
manual mode test “START”:
Rotate navigation knob (5) until voltage field is
framed. The dot in upper right hand corner
indicates that the test voltage is in preset mode. To
modify the value, rotate navigation knob (5).
•
Min. test voltage = 0.0 kV
•
Max. voltage = 24.0 kV rms (VLF),
34.0 kV (DC)
To accept the value, push in knob (5).The dot in
upper right hand disappears indicating that the test
voltage is set.
Manual Mode Setup
Waveform: Pulse
Set Pulse/Period
1:3 / 4 s
1:5 / 4 s
1:5 / 6 s
1:9 / 6 s
CANCEL
Manual Mode Setup
Waveform Sine 0.1Hz
Auto Frequency Adjust
Set Duration / Timer
15 min
CANCEL
OK
Manual Test Sine 0.1Hz
Timer: 15 min.
MAIN SETUP
Trip out on Arc
START
24.0
kV
rms
Manual Test Sine 0.1Hz
Timer: 15 min.
MAIN SETUP Trip out on Arc
START
24.0
kV
rms