User Manual
Rev 9.0
Page 17
HV Diagnostics
Switches and Controls, continued
Pos. Name
Description
8
Main switch
[on/off]
Activates the HVA.
This switch is a fuse with integrated magnetic auto-reset 10A
•
To reset
Turn the main switch OFF and then ON again
9
Power supply
plug
Serves as connection point from the HVA to the 110V – 240V,
50/60 Hz power source.
10
Grounding
connector
Serves as connection point from HVA to earth.
11
HV output
connector
Serves as connection point from the HVA to the HV test lead.
To connect
Screw the HV test lead into the HV output connector
and tighten
12
Communication
port
Serves as connection point from the HVA to PC (via RS232) or to
a USB device (via USB Flash adapter).
13
Remote control
interlock plug
Provides interlock for the remote switch (i.e. door interlock).
Can be connected to a remote emergency off switch, a gate, foot
pedal or dead man switch
14
Air Vent
Air inlet with filter, for cooling of electronic elements.
15
Air Vent
Air outlet, for cooling of electronic elements.
Right side
Left side
Detail Pos. 12
RS232
USB Adapter
11
8
10
12
13
14
15
9
Содержание HVA 34
Страница 2: ...User Manual HV Diagnostics Page 2 Rev 9 0...
Страница 48: ...User Manual HV Diagnostics Page 48 Rev 9 0...