Huvitz HRK-8000A Скачать руководство пользователя страница 58

---------------------------------------------------------------------------------------------- Auto Ref/Keratometer HRK-8000A    57

 

 

 

 

Phenomenon3: Key or JoySticik does not work

Take off the cover

Check the cable of mainboard with Key or Joystick

((When pressing the button, check whether the voltage 

goes to 'L' level. Default is 'H' level))

Does the level goes to 'L'(0V) ?

Yes

Replace Main Board

No

Check the status of Key or Joystick 

cable

(Almost of this case, the cable 

was unconnected)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Содержание HRK-8000A

Страница 1: ...Service Manual Auto Ref Keratometer HRK 8000A...

Страница 2: ...publication has been carefully checked and is believed to be entirely accurate at the time of publication HUVITZ assumes no responsibility however for possible errors or omissions or for any conseque...

Страница 3: ...age less than 250Vac minimum 6A Type SJT or SVT 18 3AWG 10A max 3 0m long One end terminates with blade attachment plug HAR Type NEMA 6 15P This product may malfunction due to electromagnetic waves ca...

Страница 4: ...ls 6 3 Environment Factors 7 4 Safety Precautions 9 III Product Outline 11 1 Nomenclature 11 2 Optical System 13 3 Electrical System 16 4 Mode Setting 24 5 Principles of Measurement 25 IV Repair Stand...

Страница 5: ...he information in this publication has been carefully checked and is believed to be entirely accurate at the time of publication Huvitz assumes no responsibility however for possible errors or omissio...

Страница 6: ...operator It is for this reason that the following safety notices have been placed appropriately within the text of this manual to highlight safety related information or information requiring special...

Страница 7: ...uired to do the protective ground to the conductive section in the equipment of class I for the purpose of safety UL60601 1 CAN CSA C22 2 NO 601 1 MEDICAL EUIPMENT WITH RESPECT TO ELECTRIC SHOCK FIRE...

Страница 8: ...requently Normal temperature for operation of the machine is at the range of 10 C 35 C and the humidity is at the range of 30 70 Do not disassemble or open the machine The manufacture shall have no re...

Страница 9: ...1060hpa For the Transformation of the machine please keep the ambient temperature is 40 70 humidity is 10 95 and atmospheric pressure is 500 1060hpa For the Storage of the machine please keep the amb...

Страница 10: ...er splashes drips or sprays Do not place containers containing fluids liquids or gases on top of any electrical equipment or devices 3 The equipment must be operated only by or under direct supervisio...

Страница 11: ...ate for the receptacle socket If you cause any damage to a cable connector s or receptacle s let the damage s be repaired by an authorized service technician 12 Please do not pull on any cable Always...

Страница 12: ...ement Button Performing the measurement by pressing it after focusing 3 Operation Lamp Indicates whether or not the electric power is on 4 Display Monitor Monitor for measurement 5 Chinrest up down bu...

Страница 13: ...connecting to the exterior equipment 3 Exterior Monitor Connection Connector Connecting into the exterior monitor 4 Clamping Bolt Fixing the system stage 5 HDMI Connector Connecting into the HDMI cab...

Страница 14: ...Auto Ref Keratometer HRK 8000A 13 3 2 Optical System...

Страница 15: ...ction mirror 6 polarizing beam splitter 7 reflection mirror 8 polarizing filter 9 collimation lens 10 diffuser 11 light source R 12 objective lens 13 insert lens high diopter 14 image lens 15 MLA micr...

Страница 16: ...light source 11 10 9 8 7 6 5 1 1 Projection system of the REF measurement 1 5 1 6 12 13 14 15 16 Illumination system of the light source for Retro Illume observation 11 10 9 8 7 6 5 1 1 Image capturi...

Страница 17: ...stem functional block diagram AC 100 240 Power switch DPST SMPS DC Output 5V 4A 5V 2A 12V 1 8A Dual Line Fuse SMPS Output DC12V 8 3A MOTER INTERFACE BOARD JP1 JP7 JP9 JP3 JP2 JP10 JP8 JP5 JP4 DC12V DC...

Страница 18: ...Auto Ref Keratometer HRK 8000A 17 3 3 1 2 Power input electrical block diagram AC 100 240 Power switch DPST SMPS DC Output 5V 4A 5V 2A 12V 1 8A Dual Line Fuse SMPS Output DC12V 8 3A...

Страница 19: ...18 Auto Ref Keratometer HRK 8000A 3 3 2 Electrical wiring 3 3 2 1 Main Board FPGA CPU FLASH SRAM SRAM SRAM SRAM SDRAM 1 2 9 3 8 7 6 5 4 10 11 12 13 14 15 16 17 dd dd d 18 19 20 21 22 23...

Страница 20: ...CN4 EXT LED EXT ILM assy 9 CN13 Mire Ring LED Mire Ring Board CON1 10 CN9 Peripheral LED Peripheral assy 11 CN10 Main Power Base Interface Board CN24 12 CN23 RGB Out Base Interface Board CN8 13 CN31 H...

Страница 21: ...20 Auto Ref Keratometer HRK 8000A 3 3 2 3 Base IO Interface Board 1 19 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 20 21 22 23 24 25 26...

Страница 22: ...ard ass y 10 CN24 Power to Main Board Mainboard CN10 11 CN10 UART Interface 2 channel Extout Board 12 CN12 Printer Switch Input Printer Button Board CN1 13 CN18 RL Encoder Input Encoder Board ass y 14...

Страница 23: ...Main assy CN2 B JP9 Motor board power Main assy CN11 C JP7 DC motor DC motor assy D JP3 Hi diopter motor diopter motor E JP2 REF PI chart lamp PI assy J2 chart lamp assy con1 F JP10 Ref LED Aiming VR...

Страница 24: ...f Main Board ID Ref No Part name Description 1 U1 S3C2410 Samsung ARM CPU 2 U33 A3P1000 FG484 Actel FPGA Logic Array 3 U31 SAA76113 Video Decoder 4 U37 MDIN 241 Deinterlacer with HDMI Out 5 U3 39VF640...

Страница 25: ...ntent of table Start Key Description Printer Button S W Upgrade Mode Download OS program into ROM via serial port ex COM permanently After downloading always HRK 7000A will start with new OS program J...

Страница 26: ...he examinee s eye is emmetropia hyperopia or emmetropia the reflected light from the retina is parallel to the optic axis of the eye diverge or converge respectively 3 The light beam reflected from th...

Страница 27: ...26 Auto Ref Keratometer HRK 8000A Objective Lens Image Lens MLA Micro Lens Array TV Camera...

Страница 28: ...inor axis 2b and then the rotational angle axis on the minor meridian Furthermore there is a correlation between the radius of cornea R the height of the projected ring image H which can be obtained f...

Страница 29: ...28 Auto Ref Keratometer HRK 8000A Ring type light source for measuring Mire Ring TV Camera Lens for measuring...

Страница 30: ...itreous humor and some of it go through the crystalline lens and the iris If it is captured by TV camera any turbid part of the transparent vitreous humor lens and cornea can be observed 3 5 3 2 Usage...

Страница 31: ...wait several seconds until the measure screen in REF mode is displayed Press the button and touch screen in following orders 1 Press Printer button 2 Touch the right of the screen of the area of meas...

Страница 32: ...output 4 Stored in FRAM Setup To import data LOAD DATA Press When loading is complete continuous beep sound is output 4 1 3 REF data set REF Test Mode A Start REF Test Mode in the Engineer mode B The...

Страница 33: ...to the delimited area shown in yellow 9 INCR After the capture when applied Threshold increases the level of Threshold In R LED camera mode Increases the brightness of REF LED illumination 10 DECR Aft...

Страница 34: ...and check the new video that is applied new level F Check the measurements of REF in the following ways 1 Select the K CCD camera mode focus on the model eye 2 Select the R CCD camera mode 3 Press bu...

Страница 35: ...ild is 1 DIOPT diopter of model eye 2 GX average horizontal distance of REF Spots 3 GY average vertical spacing of REF Spots 4 SPH Spherical coefficients of the Zernike 5 CYL Cylinder coefficients of...

Страница 36: ...n the screen Continuous press K CCD ALIGN CLIP mode switches 2 ALIGN Guide lines for the alignment of the mire ring and the LED signal displayed on the screen 3 CLIP Output calculated KER data in real...

Страница 37: ...ck to set up Focus 7 In the same way of 6 all five steps to set up a Focus 8 Upon completion of the entire Focus setup setup of the model eye proceeds KER D KER in the following ways to proceed to the...

Страница 38: ...s engineer mode is stored in the SAVE DATA 4 1 6 Check the setup data KER KER Setup Data A Begin KER Setup Data in the engineer mode B Verify the data of KER model eye 1 DIOPT Diopter of the Model eye...

Страница 39: ...2 ALIGN Select the motor to move to the center position 3 FOCUS Auto tracking begins 4 FCW Select a clockwise direction to the motor Clock Wise direction to drive a period of time 5 FCCW Select a cou...

Страница 40: ...wing ways to perform the sort 1 X HEAD Y BASE Z HEAD select the item ON 2 ALIGN button select the drive axis to move to the center position F The following is an automated tracking tests 1 Mount the m...

Страница 41: ...or of the optical base section 3 Z HEAD The optical head portion of the Z axis drive motor C For each of the motors are the following configurable items 1 BASE DUTY Motor Power of main drive 2 MOVE LU...

Страница 42: ...otor drive on the offset of the center of the range of values D For each set of items set the left and right side of the value or click the plus sign E After the Motor Setup engineer mode is stored by...

Страница 43: ...ing Slit pitch Offset increases 4 STEP The offset of the slit spacing Slit pitch Offset decreases 5 EXIT PD Setup will exit Slit pitch Offset is changed the screen update interval of the slit is measu...

Страница 44: ...ate SAVE button and press Save If you are not adequate adjust the offset of the slit spacing and try again seen D After PD Setup engineer mode by pressing the SAVE DATA is stored 4 1 10 Measuring Pref...

Страница 45: ...is the range of pixels in the direction of Focus 8 FOCUS YA Y direction pixel Focus Range 9 FOCUS ZF Z direction of the front side pigeoman direction and the pixel range of Focus 10 FOCUS ZR Z direct...

Страница 46: ...DPC function is off 3 CAMERA KER CCD Camera Color CCD and is one of the select Selected camera is displayed on the screen of the kind 4 COLOR Dead Pixel color is selected as one of the White and Blac...

Страница 47: ...ER CCD s Black Pixel in 3 COLOR CCD s White Pixel 4 COLOR CCD s Black Pixel in E After DPC Setup engineer mode by pressing the SAVE DATA is stored F HRK 8000A restarts and were being corrected by the...

Страница 48: ...ram will show the following screen 6 Select Serial Port Transmit menu then File Dialog box pops up 7 Choose HRK_test bin file then press OK button 8 After downloading turn off and on the HRK 8000A 9 C...

Страница 49: ...ews 4 of the Main body 9 and then separate the Head Rest Assy 11 2 Remove the BOLTCAP 5 10 the screws 2 and then separate top cover 6 3 Remove the screws and then separate left cover 8 and right cover...

Страница 50: ...and then separate the m front case assy 6 6 Remove the screws 1 and then separate the m middle case assy 7 7 Remove the screws 1 and then separate the m back case assy 5 8 Remove the screws 2 and the...

Страница 51: ...then remove the LCD ASSY 6 Remove the screws and then remove the OS ASSY 8 Remove the screws and then remove the XZ STAGE ASSY 9 Remove the screws and then remove the IF PCB BRACKET ASSY 5 Remove the...

Страница 52: ...en remove the lcd hinge assy L R 14 15 Remove the screws 1 and then remove the lcd back case 8 Remove the screws 5 and then remove the lcd guide 2 4 Remove the screws and then remove the lcd bracket 3...

Страница 53: ...then separate the MAIN PCB 7 Remove the screws and then remove the LCD GUIDE 4 Remove the screws and then remove the OS FRONT FRAME 2 Remove the screws and then remove the OS BACK FRAME 3 Remove the s...

Страница 54: ...COVER 12 Remove the screws and then separate the BASE COVER ASSY 13 Remove the screws and then separate the SMPS ASSY 16 Remove the screws and then separate the CONNECTOR MAIN BRKT 14 Remove the screw...

Страница 55: ...reference ring is not displayed The aiming spot is not displayed The anterior part of eye is not displayed 3 Troubles related with measurement NO TARGET is displayed OVER SPH is displayed OVER CYL is...

Страница 56: ...stered well 2 Check the voltage 5V of connector by using DMM Please refer to 3 3 section of Service Manual for Electrical Wiring Voltage 5V OK Take off the head Top cover Replace Main Board In this ca...

Страница 57: ...LCD Monitor 1 Main Board to LCD Moniter 2 Interface Board to LCD Inverter Board Please refer to the Service Manual for Electrical Wiring Does all connections are OK Yes Check voltage of the cable fro...

Страница 58: ...the cover Check the cable of mainboard with Key or Joystick When pressing the button check whether the voltage goes to L level Default is H level Does the level goes to L 0V Yes Replace Main Board No...

Страница 59: ...ing result is OK Yes Remove printer cover from the body front cover Replace the thermal printer module through hole at right side of body front cover Unconnect the cable from IO Board to the thermal p...

Страница 60: ...y of TEST MODEL EYE is very different from the original value when the focusing bright spot is somewhat out focused Enter into REF TEST Enter into TEST MODE Enter into CCD mode by pressing CCD button...

Страница 61: ...rease DIST CONST of the direction Moving too slow around focus Moving too fast or Passing away the focus Increase PWM BASE of the direction 1 or 2 steps Increase DIST CONST of the direction Decrease P...

Страница 62: ...RK 8000A 61 Phenomenon 9 Chin rest movement is not working Press Up or Down button for Chin rest movement in measure mode Not moving on any direction Yes No Okay Check the height level of chin rest Co...

Страница 63: ...to the eyesight of examinee Foggy method is the method that the size of the target is modified to be greater than the eyesight of the examinee by 1D 2D in order to reduce the error caused by this acco...

Страница 64: ...according to examinee in User Setup Mode VD 0 0mm hard contact lens VD 12 0mm lens of glasses oriental people VD 13 5mm lens of glasses western people VD 15 0mm lens of glasses western people 3 Cylind...

Страница 65: ...RK 8000A and minor meridian is displayed as CYL MIX sign will be attached in the case of the myopic astigmatism whereas sign in the case of the hyperoptic astigmatism The above mentioned option can be...

Страница 66: ...in the figure below 5 Pupil Distance PD Pupillary distance PD is the distance between the centers of the lenses of glasses Conveniently it is said that the distance between the pupils is the distance...

Страница 67: ...of the SETUP MODE See the following table and diagram for further information Classification of Data NORMAL MYOPIA HYPER OPIA ASTI GMA TISM NORMAL 0 5 S 0 5 0 5 C 0 0 MYOPIA S 0 5 0 5 C 0 0 SIMPLE MYO...

Страница 68: ...RMAL e MIXED ASTIGMATISM b MYOPIA NEAR SIGHTEDNESS f HYPEROPIA FAR SIGHTEDNESS c SIMPLE MYOPIC ASTIGMATISM g SIMPLE HYPEROPIC ASTIGMATISM d COMPOUND MYOPIC ASTIGMATISM h COMPOUND HYPEROPIC ASTIGMATISM...

Отзывы: