56 Auto Ref/Keratometer HRK-8000A ------------------------------------------------------------------------------------------------
Phenomenon2: LCD monitor display nothing(But ohter functions works)
Check whether Brightness Knob is turned well
Take off the head Top cover
Check the cables related to LCD Monitor
1. Main Board to LCD Moniter
2. Interface Board to LCD Inverter Board
(*Please refer to the Service Manual for Electrical Wiring)
Does all connections are OK?
Yes
Check voltage of the cable from Interface Board to
LCD Inverter Board is +5V
Voltage is correct?
Replace the SMPS
Yes
Replace the LCD Monitor Module
No
Make correct connection
No
External Monitor is OK?
Yes
Replace the Main Board
No