Getting started instructions for Tracker 3200S and 4.3 Software
21-2615 Version 1.1
2
Do not run the software at this point. The hardware must now be installed to
complete the installation.
1-3 CONNECTING THE TRACKER 3200S
Note for Windows 7 64bit users:
The device driver may attempt to auto install and fail.
After following the steps below (they may not all occur), open device manager and look
under Huntron devices for the Tracker 3200S with a yellow exclamation point. Update this
driver by right clicking on them. Select the install DVD for the source of the driver if
needed.
1. Connect the Tracker 3200S power cord to a power outlet.
2. Connect the USB cable to the Tracker 3200S and the PC.
3. Turn on the Tracker 3200S power
4. An
Add New Hardware
wizard prompts you to install a
“Huntron Tracker Model
3200S
”.
If prompted to find driver on the Internet select No and Click Next.
5. If the wizard prompts you to automatically find drivers, click
Next.
6. When a message appears, indicating that the driver has not passed Windows Logo
testing, click
“Continue anyway”.
Then click
Finish.
7. Install any other equipment to be used with the Huntron Workstation software.
8. Remove your Huntron Workstation DVD. You are now ready to use the software.
9. A Huntron Workstation icon was created on your windows desktop. Double click
this icon to start the software.
10. The Huntron Workstation Activation dialog is displayed. This software is only
licensed to be used on one Tracker 3200S. The activation codes are generated by
Huntron and are specific to the Tracker 3200S serial number. Activation codes are
printed on the software DVD. If you do not know the activation codes, call Huntron
technical support (425-743-3171) and provide the serial number for the Tracker
3200S. Tracker 3200S serial numbers start with a
32S.
1-4
CONFIGURING THE TRACKER 3200S
1.
When the Workstation main screen is displayed, on the main menu, select Tools-
>Options->Hardware. Set the Tracker to Tracker 3200S. Set the Tracker Port to
USB.
2.
Set the Prober settings as needed (refer to Prober Getting Started document).
3.
Click the connect buttons for the Tracker and Prober if used.
4.
Click the Save button.
5.
NOTE: The Tracker 3200S is already calibrated at the factory. However, if you are
connecting a Tracker 3200S to a Prober, you will need to calibrate it as a complete
system.
6.
Connect the Tracker 3200S to your Prober, follow the steps below to calibrate the
system.
7.
Tracker 3200S Calibration for Prober
To access the calibration window, click
Tools->Maintenance->Tracker3200S
Diagnostics
.