REPLACINg the SENSOR
•
Carefully remove the sensor housing cover.
•
Always ‘gently’ rotate (quarter turn) the sensor housing cover
while fitting or removing.
•
Remove the sensor using the sensor removal tool by hooking
it into the groove on the black part of the sensor, then simply
rest the removal tool against the monitor and lever the sensor
gently upwards.
•
There is no membrane cover to remove on the sensor.
•
After installation of the new sensor, reset the monitor by pressing
the power button and calibration button at the same time then
carry out the zero correction facility.
Installing a New Sensor
Note: No force should be applied to the sensor, the sensor housing or the sensor
cover. Do not squeeze the sensor with the sensor removal tool.