SpectraTrend HT User’s Guide 3.2
66
Figure 74. Reading the Black Glass
•
Slide the calibration box to the black indication on the side of the box. The box will
rest in the first detent position
.
Press the button (
GO
). The SpectraTrend HT reads
the black glass and sets the instrument zero. When it is finished, the screen
prompts you to place the white tile.
Figure 75.Reading the White Tile
•
Replace the black glass by sliding the white tile into position, which is the second
indentation on the standards box.
•
Press the button (
GO
). The SpectraTrend HT reads the white tile and sets the top of
scale.
•
Replace the white tile by sliding the green tile into position, which is the third
indentation on the standards box.