11/02
Specifications
5-5
Conditions of Illumination and Viewing
Geometry:
0°/45° circumferential viewing.
Viewed area:
2.5” (65 mm) diameter (LAV), 1.25” (30 mm) diameter (SAV).
Light source:
Xenon flash
Source UV content:
Excess UV vs. D65 with UV control option
Monochromator:
Diffraction grating/detector array
Port orientation:
Standard:
Horizontal sample plane, port down
Optional:
Horizontal sample plane, port up
Vertical sample plane, port sideways
Source lifetime:
>1.5 billion flashes (>6 months of continuous operation at the fastest
measurement speed, 100 Hz).
Standoff distance:
1 inch
Standoff tolerance (for
maximum allowable
change in L* of 0.5):
±
1/4 inch
Measurement speed: 100 Hz.