version 1.2_en
page 17
Calibration curves conductance sensor
calibration curve
description
density
measuring
range
brick
1.0 to 30%
gypsum
0.5 to 14%
lime mortar
0.5 to 10%
lime gypsum plaster
approx.
1150kg/m³
0.1 to 10%
lime cement plaster
approx.
1250kg/m³
0.4 to 10%
lime plaster
approx.
1450kg/m³
0.8 to 10%
cork
4.0 to 30%
cellulose
injected insulation
material
approx.
55kg/m³
10.0 to 40%
Digit
-------
0 to 100
test block
! Only for checking the
instrument !
-------
-------
Calibration curves for plaster
Due to the multitude of diverse plastering mixtures, the calibration curves
have been determined based on a selection of various plastering mixtures of
different producers, which have been summarised in the above mentioned
main groups of plaster.
The curves have been determined in plastered condition. It is not possible to
measure the powder or not processed plaster!