the menu. 
NOTE: If the device exceeds the full scale value, the result will start 

flashing. 
It is enough to calibrate a single unit for all other units except Bar unit. 

The device will translate this calibration for other units. 
3. BUTTONS
ZERO: When this button is pressed, the device output is reset. This 

button has two different functions as "Up-arrow" and "Zero". The up-

arrow function is used when changing options in the menu. The "Zero" 

function is only required during calibration. "Unloaded state" is required 

when calibrating. When the unloaded state is defined for the device, the 

user must apply pressure slowly to a load cell to find the nearest point to 

the zero point. At this pressure point the user presses the "zero" button 

of both the device and the load cell and defines this pressure point as 

zero for both. This is some kind of fine tuning. 
Unloaded state: No pressure load on the device. 
PEAK: When this button is pressed, the device will show the maximum 

pressure point at that time. For example; Assume that we make a 

pressure measurement as in Figure 1 and press the peak button at 

different times. 
First time interval (A-B): The measurement results are increasing in the 

first time interval. So, whichever point in this region is pressed on the 

peak button, our maximum measurement will be the last measurement 

we made. However, the result at point B has begun to decrease. 
Second time interval (B-C): Starting from point B until the point at which 

the pressure has passed, the pressure value at the maximum measured 

value B in this range will be. 
Third time interval (C-D): The area between C-D is the same as the 

whole procedure A-B. By the way, our maximum measurement will be 

the last measurement we made. 
Fourth time interval (D-E): The pressure will remain at the maximum 

measured value D at the pressure point D until the last time.

Содержание HS-4157

Страница 1: ...product manual 06 18 HS 4157 Point Load Tester ...

Страница 2: ......

Страница 3: ...ds 2 1 Uni X Units The user can view the instrument s measurement with 5 different units The SET button must be pressed to change the unit After viewing Uni X on the screen it can be easily changed with the up arrow button All units are given below together with unit numbers Uni 0 bar Uni 1 kg Uni 2 tone Uni 3 Newton Uni 4 kN 2 2 FL XX Filter The filter number is the number of samples that the dev...

Страница 4: ...vice will show 20 of the full scale In this step a reliable load cell must be applied to the device by measuring the pressure load cell that the user requests When the correct pressure is reached press the zero button to continue This procedure will be repeated 6 times in total The pressure points are as follows 1 0XX 20 2 0XX 40 3 0XX 60 4 0XX 80 5 0XX 100 full scale After pressing the zero butto...

Страница 5: ... This is some kind of fine tuning Unloaded state No pressure load on the device PEAK When this button is pressed the device will show the maximum pressure point at that time For example Assume that we make a pressure measurement as in Figure 1 and press the peak button at different times First time interval A B The measurement results are increasing in the first time interval So whichever point in...

Страница 6: ...5 ON SET This button is used to switch the device ON and OFF and also it is used to change the menu ...

Страница 7: ...7 ...

Страница 8: ...Humboldt Mfg Co 875 Tollgate Road Elgin Illinois 60123 U S A U S A Toll Free 1 800 544 7220 Voice 1 708 456 6300 Fax 1 708 456 0137 Email hmc humboldtmfg com ...

Отзывы: