INTRODUCTION
The HS-4157 Point Load Tester is a high-resolution digital pressure 

gauge that measures with high accuracy. Easy to install and critical 

pressure measurement is designed to be able to make a high 

measurement in the field. Two standard AA alkaline batteries can run 

continuously for more than 10 months. 
1. BEFORE SET UP
Please read all instructions carefully before setting up the display. If 

there is an error in the installation, the indicator may be permanently 

damaged. The installation should be carried out by qualified personnel 

who are familiar with the standards for safety practices and pressure 

systems. 
2. HOW TO USE MENU
Press the "ON" button once to turn the device on. To switch off the 

device, hold down the "ON" button for 3 seconds. 
2.1.  Uni X (Units) 
The user can view the instrument's measurement with 5 different units. 

The "SET" button must be pressed to change the unit. After viewing 

"Uni X" on the screen, it can be easily changed with the up arrow 

button. All units are given below together with unit numbers:
Uni 0: bar 
Uni 1: kg 
Uni 2: tone 
Uni 3: Newton 
Uni 4: kN 
2.2. FL XX (Filter) 
The filter number is the number of samples that the device has received 

at intervals of 100 ms, and the device will display the measurement 

result by taking the average of these samples. For example, select "FL-

02" as the filter number. The device will take 2 measurements in 100 ms 

interval and take the average of them and calculate the measurement 

result. The filter number can be selected from 1-99.
NOTE: If "FL 00" is selected as the filter number, the device can not 

perform any filtering and measurement. The result starts to flash at 

a fixed number. So this device should not be operated with "FL 00" 

option. 

2

Содержание HS-4157

Страница 1: ...product manual 06 18 HS 4157 Point Load Tester ...

Страница 2: ......

Страница 3: ...ds 2 1 Uni X Units The user can view the instrument s measurement with 5 different units The SET button must be pressed to change the unit After viewing Uni X on the screen it can be easily changed with the up arrow button All units are given below together with unit numbers Uni 0 bar Uni 1 kg Uni 2 tone Uni 3 Newton Uni 4 kN 2 2 FL XX Filter The filter number is the number of samples that the dev...

Страница 4: ...vice will show 20 of the full scale In this step a reliable load cell must be applied to the device by measuring the pressure load cell that the user requests When the correct pressure is reached press the zero button to continue This procedure will be repeated 6 times in total The pressure points are as follows 1 0XX 20 2 0XX 40 3 0XX 60 4 0XX 80 5 0XX 100 full scale After pressing the zero butto...

Страница 5: ... This is some kind of fine tuning Unloaded state No pressure load on the device PEAK When this button is pressed the device will show the maximum pressure point at that time For example Assume that we make a pressure measurement as in Figure 1 and press the peak button at different times First time interval A B The measurement results are increasing in the first time interval So whichever point in...

Страница 6: ...5 ON SET This button is used to switch the device ON and OFF and also it is used to change the menu ...

Страница 7: ...7 ...

Страница 8: ...Humboldt Mfg Co 875 Tollgate Road Elgin Illinois 60123 U S A U S A Toll Free 1 800 544 7220 Voice 1 708 456 6300 Fax 1 708 456 0137 Email hmc humboldtmfg com ...

Отзывы: