06.18

 HS-4157

Point Load Tester

Содержание HS-4157

Страница 1: ...product manual 06 18 HS 4157 Point Load Tester ...

Страница 2: ......

Страница 3: ...ds 2 1 Uni X Units The user can view the instrument s measurement with 5 different units The SET button must be pressed to change the unit After viewing Uni X on the screen it can be easily changed with the up arrow button All units are given below together with unit numbers Uni 0 bar Uni 1 kg Uni 2 tone Uni 3 Newton Uni 4 kN 2 2 FL XX Filter The filter number is the number of samples that the dev...

Страница 4: ...vice will show 20 of the full scale In this step a reliable load cell must be applied to the device by measuring the pressure load cell that the user requests When the correct pressure is reached press the zero button to continue This procedure will be repeated 6 times in total The pressure points are as follows 1 0XX 20 2 0XX 40 3 0XX 60 4 0XX 80 5 0XX 100 full scale After pressing the zero butto...

Страница 5: ... This is some kind of fine tuning Unloaded state No pressure load on the device PEAK When this button is pressed the device will show the maximum pressure point at that time For example Assume that we make a pressure measurement as in Figure 1 and press the peak button at different times First time interval A B The measurement results are increasing in the first time interval So whichever point in...

Страница 6: ...5 ON SET This button is used to switch the device ON and OFF and also it is used to change the menu ...

Страница 7: ...7 ...