background image

2

While facing the back of the panel, the four connections starting from 

the left are for the water drain, tap water input, vacuum input, and 

compressed air input respectively. The last three connections are for  

de-aerated water return from the de-airing tank, tap water going to the 

de-airing tank, and air/vacuum to the de-airing tank respectively.
On the face of the panel are the connections for air pressure outputs to 

the HM-2450A Automatic Pressure Controllers, tap water and  

de-aerated water for the HM-4151A Air/Water Bladder Cylinders,  

HM-2315 Automatic Volume Change, and the triaxial cell. Also, there 

is an auxiliary pressure connection for air pressure up to 10 psi and an 

auxiliary vacuum connection. These connections are made using quick 

disconnect style fittings. These fittings have a barbed end which fits 

into the tubing and a nut which fits over the tubing to clamp the tubing 

on to the barbed end of the fitting. To disconnect the fittings from the 

panel, push down on the tab at the top of the socket half of the fitting 

as shown in Figure 3. The plug half of the fitting will pop outward.  To 

reconnect it, just push the plug into the socket until the tab clicks and 

locks the plug in place. The connection for the tap water to the base 

of the sample via the volume change device requires a ¼ inch to 1/8 

inch push-to-connect adapter to allow the attachment of ¼ inch tubing 

from the panel to the 1/8 inch OD tubing which connects to the volume 

change device. This adapter is supplied with the panel. 

BLADDER CYLINDER CONNECTION

The HM-4151A Air/Water Bladder Cylinders have push to connect 

style fittings at the bottom of the assembly. These fittings use the ¼ 

inch OD tubing and attach the by the same method as the fittings on 

the back side of the panel. The right fitting is for air pressure from the 

Automatic Pressure Controller.  The other push-to-connect fitting is for 

water from the connections on the front of the panel. There are two 

bladder cylinders required per triaxial cell. One bladder cylinder is for 

cell pressure and tap water supply, the other cylinder is for back pressure 

and de-aerated sample water supply. The air pressure line for one of the 

bladder cylinders will connect to the Cell air pressure connection on the 

back of the HM-2450A Automatic Pressure Controller and the water line 

for that bladder cylinder will connect to the Bladder A connection on 

the front of the panel. The air line for the second bladder cylinder will 

connect to the back pressure connection on the back of the HM-2450A 

Automatic Pressure Controller and the water line for the second bladder 

cylinder will connect to the Bladder B connection on the front of the 

panel as shown in Figure 2.

AUTO VOLUME CHANGE CONNECTION

The HM-2315 Automatic Volume Change Device has push-to-connect 

fittings for 1/8 inch OD tubing. The volume change device is connected 

Содержание HM-4154

Страница 1: ...product manual HM 4155_MAN_0812 HM 4155 HM 4154 3 Cell Auto Control Panel 1 Cell Auto Control Panel ...

Страница 2: ...a simple free standing bench mounted system which requires no fixed mounting PANEL MOUNTING A pair of supporting feet is supplied with the panel so that it can be bench mounted and remain stable Screws nuts and washers for the support feet are supplied so that the support feet can be attached as shown in Figure 1 AIR WATER CONNECTIONS All connections will require either 1 4 inch OD or 1 8 inch OD ...

Страница 3: ...onnect adapter to allow the attachment of inch tubing from the panel to the 1 8 inch OD tubing which connects to the volume change device This adapter is supplied with the panel BLADDER CYLINDER CONNECTION The HM 4151A Air Water Bladder Cylinders have push to connect style fittings at the bottom of the assembly These fittings use the inch OD tubing and attach the by the same method as the fittings...

Страница 4: ...electrical connections required on the HM 4155 Distribution Panel OPERATION After making all of the necessary connections check to make sure all of the valves on the front of the panel are set to the off position Also loosen the jam nuts on the regulators as shown in Figure 3 AIR PRESSURE ADJUSTMENT There are two separate regulators on the panel The main regulator located below the gauge labeled S...

Страница 5: ...o the corresponding bladder cylinder and the other controls the flow of de aerated water to the Auto Volume Change and to the other bladder cylinder in the apparatus Figure 8 When the valve for the cell is in the FILL position water flows to the cell and bladder cylinder at the same time When the valve labeled PWP is in the FILL position de aerated water flows to the sample base through the Automa...

Страница 6: ...pressure If so the fluids will spray out at high velocity If the filter becomes dirty and clogged open the drain plug as described above to relieve pressure Remove the bowl by rotating the bowl counterclockwise to unscrew it from the filter body Be careful not to lose or damage the o ring that seals the bowl to the body of the filter Next unscrew the filter support by turning it counterclockwise u...

Страница 7: ...99B TRIAXIAL CELL HM 4170 PWP HM 3000 DIGITAL MASTER LOADER HM 2450A AUTOMATIC PRESSURE CONTROLLER HM 2315 AUTOMATIC VOLUME CHANGE DEVICE HM 4151A BLADDER AIR WATER CYLINDER HM 2310 10 DISPLACEMENT TRANSDUCER HUMBOLDT MFG CO FULLY AUTOMATED 3 TRIAXIAL CELL SYSTEM COMPRESSED AIR VACUUM TAP WATER DRAIN TO BASE VALVE TO CELL VALVE FIGURE 2 ...

Страница 8: ...R PRESSURE PUSH DOWN TO RELEASE CONNECTOR FIGURE 3 Figure 3 Figure 4 Triaxial Cell BASE WATER CONNECTION DISTRIBUTION PANEL SOLENOID VALVE PORE WATER PRESSURE TRANSDUCER V1 V2 V3 V4 V5 VENT VALVE LEVER FIGURE 4 AUTO VOLUME CHANGE CONNECTION ...

Страница 9: ...SE PRESSURE TO THE DE AIRING TANK TO THE AUXILIARY AIR CONNECTOR FIGURE 5 REGULATING AUXILIARY AIR PRESSURE FIGURE 6 DE AIRING TANK CONTROL VALVES DE AIRED WATER FILL DRAIN OFF VACUUM PRESSURE TURN KNOB LEFT TO FILL DE AIRING TANK WITH TAP WATER OR TURN KNOB RIGHT TO DRAIN DE AIRING TANK TURN KNOB LEFT TO APPLY VACUUM TO DE AIRING TANK OR TURN KNOB RIGHT TO APPLY PRESSURE TO ASSIST DRAINING OF DE ...

Страница 10: ...URE 7 AIR PRESSURE CONTROL VALVE WATER CONNECTION CELL BLADDER A FILL DRAIN BASE BLADDER B FILL CELL PWP FIGURE 8 CELL AND SAMPLE BASE WATER CONTROLS TURN KNOB UP TO FILL CELL AND BLADDER CYLINDER A WITH TAP WATER TURN KNOB DOWN TO DRAIN CELL AND BLADDER TURN KNOB DOWN TO FILL SAMPLE AND BLADDER CYLINDER B WITH DE AERATED WATER 8 ...

Страница 11: ...nstream adjust drip rate accordingly Consult oil delivery conversion chart see below Generally one drop per minute downstream for every 10 15 SCFM flow is satisfactory 25 drops per minute equals one ounce per hour volume of oil passing through the sight dome NOTE This is a constant density type lubricator which delivers a constant ratio of oil to air flow Therefore if air flow increases or decreas...

Страница 12: ...r parts of such product for the warranty of this product please refer to Humboldt Mfg Co catalog on Terms and Conditions of Sale The purchaser is responsible for the transportation charges Humboldt Mfg Co shall not be responsible under this warranty if the goods have been improperly maintained installed operated or the goods have been altered or modified so as to adversely affect the operation use...

Отзывы: