background image

MEDLINK LITE USER MANUAL

Humanscale

Содержание MedLink Lite

Страница 1: ...MEDLINK LITE USER MANUAL Humanscale ...

Страница 2: ... all drawers 21 5 5 Ejecting cassettes 23 Legal Notices 6 1 Legal Notice 24 Introduction 1 1 About Humanscale Healthcare 3 1 2 Contact Information 3 1 3 Introduction to Medlink 4 1 4 About the T7TM Point Of Care Technology Cart 4 Setting up MedLink 2 1 About the MedLink drawer system 5 2 2 Unpacking and assembly 6 2 3 Installing Shelves 6 2 4 Installing Dividers 7 2 5 Inserting and removing Casset...

Страница 3: ...life and follow the belief that if a design solves a functional problem as simply and elegantly as possible the resulting form will be timeless Humanscale Healthcare applies that same design approach to ergonomic solutions that are created to improve the caregiving experience for caregivers and patients alike Humanscale Healthcare s products include the ViewPoint series of technology wall stations...

Страница 4: ...ery while enhancing patient safety MedLink is expertly designed to reduce human error and supports a closed loop medication system 1 4 ABOUT THE T7TM POINT OF CARE TECHNOLOGY CART At the core of MedLink is T7 part of the revolutionary TouchPoint line of mobile technology cards Built with the caregiver s workflow in mind T7 addresses the challenges faced by existing medical cards by offering optima...

Страница 5: ...placed and fit into the cassettes There are 3 sizes of drawers small medium and large Shelves are used to compartmentalize cassettes depending on the required drawer layout Dividers can be placed into the drawers to create multiple compartments within a single drawer Override Keys are used to unlock and gain access to the drawers in the event of a system or power failure The keys are not used duri...

Страница 6: ...a cassette you must first insert shelves to configure your required drawer layout Cassettes maybe reconfigured at any time if the layout requirements change 1 To insert a shelf hold the cassette in place on a flat surface with one hand With the other hand insert the shelf so that the slots on sides of the shelf slide over the locating pins on the inside side walls of the cassette 2 Ensure the rear...

Страница 7: ...up but when the cutouts are positioned facing upwards it allows packaged syringes to be laid flat on top of the dividers 2 5 INSERTING AND REMOVING CASSETTES Once a cassette is configured with the required shelf layout it can be inserted into the housing Slide the cassette into an available location on the housing and push it fully into place Push the cassette handle into the housing until it lock...

Страница 8: ... storage drawer for non patient specific supplies 1 When inserting drawers they must be tilted backwards slightly to ensure the tabs at the rear of the drawer clear the underside of the shelf 2 Once the back of the drawer is inserted the drawer can be tilted back to a horizontal angle and pushed all the way in When the drawer is fully inserted you should hear a click This is the locking mechanism ...

Страница 9: ... new T7 carts the default PIN code is set to 1234 You should change the PIN to a unique code that is only known to supervisors managers and technical support staff within your facility Instructions for this can be found within the T7 User Guide Initial Software Configuration The MedLink Lite control software is built into the T7 cart touch screen user interface Before you can use MedLink you must ...

Страница 10: ... control to switch it to gray and disable MedLink Pro This will enable MedLink Lite 7 When switching this control on or off you will be prompted to confirm the change Touching OK will confirm the change and the cart will reboot with the new setting configured 8 Once fully booted with MedLink Lite enabled the system should automatically detect the presence of the MedLink drawer system and the MedLi...

Страница 11: ... 1 From any screen on the interface touch the gear icon in the header to enter the Settings screen 2 Scroll to the bottom of the settings list and touch Administrative Login 3 Enter the Administrator PIN code 4 Scroll to the bottom of the Admin Settings list and touch MedLink Settings 5 The configurable MedLink Settings list is displayed 3 3 CHANGING MEDLINK LITE SETTINGS In the MedLink Lite setti...

Страница 12: ...n All function can be toggled on or off by touching the switch control Only a user with administrator access can enable the Open All function Once enabled by an administrator any user can execute the Open All function When this function is enabled an additional setting is displayed which allows you to choose the drawer close timeout for when the drawers are opened using the Open All function This ...

Страница 13: ... generally speaking it is best practice to minimize the amount of time required for this feature to activate so that access to the cart is properly controlled 3 3 5 CHANGE MEDLINK PIN Use this control to change the MedLink PIN code to a unique code that is known to caregivers Touch the control to begin the process and follow the on screen prompts as follows 1 Enter the current MedLink PIN By defau...

Страница 14: ...ill open slightly Pull the drawer forwards to access the contents Tabs at the rear of the drawer prevent the drawer from sliding all the way out accidentally 2 The drawer can be fully removed by tilting the drawer backwards in exactly the same way as when inserting the drawer and continuing to pull the drawer forwards Drawer locking system explained ...

Страница 15: ...tte handles from the housing Pulling the handles outwards will allow the cassettes to be removed from the housing 3 Second use the same override key in the lock on the rear panel of the cassette Turn the key counter clockwise to unlock all the drawers in the cassette Drawer may now be removed from the cassette NOTE Once either override lock has been actuated you must turn the key back to the verti...

Страница 16: ...ttes A and C are the cassettes on the left side of the cart cassettes B and D are the cassettes on the right side of the cart 4 The first time you use MedLink and when all the drawers are unlabeled you will be presented with the following screen prompting you to select a cassette This view is known as Cassette View 5 Touching a cassette will take you to Drawer View which shows a graphical represen...

Страница 17: ...ist on the right hand portion of the display touch the Add Label icon that corresponds to the drawer you wish to label 2 You will be presented with a list of options for labelling the drawer Touch Enter drawer label 3 3 This will take you to the label entry screen Use the on screen keyboard to enter an appropriate label You might want to use the patient s last name with first initial or the patien...

Страница 18: ...S 1 Labelling storage drawers follows the same process Touch the Add Label icon for the drawer you wish to label 2 Then when presented with the list of options for labelling the drawer touch Label drawer as STORAGE 3 You will see the drawer you chose is now labelled as a storage drawer Storage drawers in the drawer list and on the cassette graphic are represented by being underlined and having upp...

Страница 19: ... 1 OPENING A PATIENT DRAWER 1 In Drawer View to open a patient drawer first select the drawer either by touching the label in the drawer list or touching on the drawer in the cassette graphic In Cassette View you may also touch a label in the drawer list 2 The selected drawer and its label will highlight in gray helping you to locate the drawer on the physical drawer system NOTE With the drawer se...

Страница 20: ...awers which bypasses the drawer selected screen Simply double tap the drawer or label in Cassette View or Drawer View This will unlock the drawer in one step and skip to this screen 4 Upon opening or removing the drawer the display will revert to Drawer View with the drawer list and cassette graphic showing a blank label to indicate there is no drawer present 5 3 2 OPENING A STORAGE DRAWER 1 Openi...

Страница 21: ...he Open button to open it 2 The storage drawer will unlock and both drawers will be highlighted blue on the cassette graphic 5 4 OPENING ALL DRAWERS When the Open All function is enabled the Open All button is displayed in Drawer View This allows the user to unlock all the drawers in that cassette at once This should only be used when emptying the cassette of medication It should not be used when ...

Страница 22: ...e labels Touch an option to continue 4 All the drawers on the chosen cassette will unlock and highlight in blue on the cassette graphic Drawers maybe removed and emptied If drawers are not opened or removed when the re lock timer reaches zero they will automatically re lock 5 If you selected an option such as Open all Erase all labels the appropriate labels will be missing when the drawers are rep...

Страница 23: ...MedLink Pro 1 In Cassette View touch the eject button in the center of the cassette graphic 2 The cassettes will highlight in blue Touch the cassette you wish to eject 3 The cassette handle will pop out of the housing allowing you to remove it Continue to eject additional cassettes if required or touch the Cancel button to return to Cassette View 4 On returning to Cassette View you will see the ej...

Страница 24: ...es to ensure the correct medication is being administered to the correct patient in the prescribed dosage and delivery method When placing medication into MedLink drawers the medication must be in its original packaging or contained within separate closed packaging All rights reserved 2016 Humanscale Corporation The Humanscale word mark and logo are registered trademarks in the United States and c...

Отзывы: