Laser successfully connected.
Laser ON
–
Turns the laser ON. If the laser is in autostart mode this is equivalent to “restart”.
Laser OFF
–
Turns the laser OFF.
Mode
–
Gives a choice of operational modes possible to choose for the laser model. For 04 series lases Constant Power
or Constant Current operation can be chosen. Only relevant choices are presented to the mode of operation chosen.
Commands
–
opens a command communications window to send commands directly to the laser controller.
Message
–
highlights important information of the laser status to the user.
Disconnect
–
allows the user to disconnect from the Cobolt Monitor
™
software in a controlled way.
NOTICE
The communication cable should not be removed when the software is in connect state. The
communication within the controller may then malfunction and this might require a power restart of the
driver. To disconnect the laser click “Disconnect” or close Cobolt Monitor™
completely. It is also possible to
disconnect by powering the laser OFF. In this case Cobolt Monitor
™
will automatically close the window for
that laser.
Clear Fault
- is displayed in the event of a fault. The user can de
al with the cause of the fault and then press “Clear
Fault” and then restart the laser by clicking “Laser ON”. Example: if the remote interlock loop is open the user must
make sure the loop is closed again before issuing a “Clear Fault” followed by “Laser On”.
More
–
an additional Cobolt Monitor
™
window will open containing more detailed information of that laser’s status.