Connecting an externally closed application (reactor)
Remove the thread covers from the unit´s circulation flow (1) and circulation return (2).
Make sure that the hose material is compatible with the thermal fluid and temperature
range being used. The appropriate size spanners can be taken from the table below. In
order that the application can be driven correctly, and that no air bubbles remain in the
system, ensure that the unit´s circulation flow (1) is attached to the lowest connection
on the application, and the unit´s circulation return (2) is attached to the highest
connection on the application.
ministat 230(w), 240(w)
ministat125(w)
Other CCs:
Please note the markings for the hose connections on the housing.