Pub. 42004-566A 

TempSense Models TS200/TS200W 

Page 8 of 9 

P:\Standard IOMs - Current Release\42004 Instr. Manuals\42004-566A.docx 

03/21 

 

Figure 11.  Temperature Superposition and Calibration 

Operation

 

Environment

 

Environmental considerations for the Models TS200/TS200W NCIT (non-contact infra-red thermometer) 
performance: 

 

Use the temperature scanning feature of the TS200/TS200W in a draft-free space that is out of direct 
sunlight and away from radiant heat sources. 

 

The optimal environment for the TS200/TS200W temperature sensor is between 60.8–104 ºF (16–40 
ºC) with the relative humidity below 85 percent. 

 

Allow the TS200/TS200W to operate in the testing environment for 10–30 minutes to adjust to its 
environment before scanning personnel. 

Subject Preparation

 

To obtain accurate skin temperatures, ensure that the subject has not: 

 

been wearing a hat, scarf, heavy clothing, etc. 

 

underwent recent physical exertion 

 

just entered from a much lower temperature environment 

 

recently used facial cleaning products 

 

engaged in any other act that may affect their skin temperature. 

Scanning Subjects

 

When a subject with a temperature exceeding the configured 

threshold

 enters the IR camera's FOV: 

 

The TempSense TS200/TS200W sends an email to the SMTP 

recipient

 containing still images of the 

IR and standard camera's FOV. 

 

The TS200/TS200W image includes the 

temperature

 ±

Offset

 superimposed in the IR picture. 

Содержание GAI-TRONICS TempSense TS200

Страница 1: ...WWW GAI TRONICS COM FOR PRODUCT LITERATURE AND MANUALS Confidentiality Notice 1 General Information 1 Installation 2 Camera Location 2 Wall Mount 3 Pole Mount 3 VESA Mount 3 Stand Mount 4 Connection...

Страница 2: ...eration of your GAI Tronics product or system This manual may not be disclosed in any form in whole or in part directly or indirectly to any third party General Information The TempSense Models TS200...

Страница 3: ...closer to the camera Camera Location Locate the TempSense TS200 TS200W so that people in a queue individually pass through the two cameras shared FOV while facing the cameras Each subject scanned shou...

Страница 4: ...mounting plate using the four included 6 32 screws PoleMount Mount the TempSense TS200 TS200W to a pole see Figure 5 as follows 1 Secure the TS200 see Figure 6 or TS200W see Figure 7 mounting plate t...

Страница 5: ...If the IR sensor LED does not flash press the sensor into its socket using a soft cloth then reboot the device The Model TS200 boots up and is ready for configuration see the Configuration section or...

Страница 6: ...and time see Table 4 5 Configure all settings and click APPLY CHANGES AND RESTART The unit reboots with the configured settings NOTE If the network mode is DHCP but no DHCP server exists the unit fal...

Страница 7: ...rts after the measured temperature drops below the threshold temperature 0 100 5 seconds Offset The value to add to or subtract from the temperature measured by the TempSense 10 to 10 F 0 F 6 to 6 C 0...

Страница 8: ...Pv4 address or FQDN string 0 0 0 0 time_zone The offset from UTC Select the offset from UTC using the roll box control Calibration 1 Place an object of known temperature in the field of view of the TS...

Страница 9: ...rcent Allow the TS200 TS200W to operate in the testing environment for 10 30 minutes to adjust to its environment before scanning personnel Subject Preparation To obtain accurate skin temperatures ens...

Страница 10: ...andard IOMs Current Release 42004 Instr Manuals 42004 566A docx 03 21 Specifications Electrical Power TS200 PoE TS200W 120 V ac Mechanical Dimensions 7 26 H 5 76 W 1 5 D in 184 4 146 3 38 1 mm Weight...

Отзывы: