Pub. 42004-566A
TempSense Models TS200/TS200W
Page 2 of 9
P:\Standard IOMs - Current Release\42004 Instr. Manuals\42004-566A.docx
03/21
Installation
N
OTE
:
The accuracy of the TempSense TS200/TS200W requires individual subjects pass within two to
six feet of the thermal imaging camera lens. Results with different camera positions may vary.
Thermal imaging camera accuracy improves as the subject moves closer to the camera.
Camera Location
Locate the TempSense TS200/TS200W so that people in a queue individually pass through the two
cameras’ shared FOV while facing the cameras. Each subject scanned should come within two to six feet
of the cameras. The ideal height for camera installation is 5 feet 5 inches above the floor. Installation at a
higher location requires aiming the TS200/TS200W downward so that each subject’s forehead passes
through the center of the FOV (see Figure 2 and Figure 3).
Figure 2. SAMPLE FIGURE Camera Positioning (Side View)
Figure 3. SAMPLE FIGURE Camera Positioning (Top View)