Subject Preparation
To obtain accurate skin temperatures, ensure that the subject has not:
•
been wearing a hat, scarf, heavy clothing, etc.
•
underwent recent physical exertion
•
just entered from a much lower temperature environment
•
recently used facial cleaning products
•
engaged in any other act that may affect their skin temperature.
Calibration
1.
Place an object of known temperature in the field of view of the TS100 at the same height and
distance from where the TS100 will scan subjects.
The calibration object must be the hottest object in the FOV.
2.
While watching the temperature of the object in the IR video feed (left side), press the setup button.
3.
Slide the OFFSET slider to adjust the on-screen temperature until it displays the known temperature
(see Figure 9).
Figure 9. Temperature Superposition and Calibration
4.
Select
F
AHRENHEIT
or
C
ELSIUS
from the
U
NITS
field on screen.
Selection of either scale reboots the TS100. The TS100 displays the temperature in the selected units
after rebooting.
5.
Adjust the
T
EMP
T
HRESHOLD
to the temperature for the TS100 to alert the operator and capture a
still photo.
The TS100 captures both an IR and normal picture when the temperature exceeds the
T
EMP
T
HRESHOLD
value.