Subject Preparation

 

To obtain accurate skin temperatures, ensure that the subject has not: 

 

been wearing a hat, scarf, heavy clothing, etc. 

 

underwent recent physical exertion 

 

just entered from a much lower temperature environment 

 

recently used facial cleaning products 

 

engaged in any other act that may affect their skin temperature. 

Calibration

 

1.

 

Place an object of known temperature in the field of view of the TS100 at the same height and 
distance from where the TS100 will scan subjects. 

The calibration object must be the hottest object in the FOV. 

2.

 

While watching the temperature of the object in the IR video feed (left side), press the setup button. 

3.

 

Slide the OFFSET slider to adjust the on-screen temperature until it displays the known temperature 
(see Figure 9)

 

Figure 9.  Temperature Superposition and Calibration 

4.

 

Select 

F

AHRENHEIT

 or 

C

ELSIUS

 from the 

U

NITS

 field on screen. 

Selection of either scale reboots the TS100.  The TS100 displays the temperature in the selected units 
after rebooting. 

5.

 

Adjust the 

T

EMP 

T

HRESHOLD

 to the temperature for the TS100 to alert the operator and capture a 

still photo. 

The TS100 captures both an IR and normal picture when the temperature exceeds the 

T

EMP 

T

HRESHOLD

 value. 

Содержание GAI-TRONICS TempSense TS100

Страница 1: ...96 5954 VISIT WWW GAI TRONICS COM FOR PRODUCT LITERATURE AND MANUALS Confidentiality Notice 1 General Information 1 Installation 2 Camera Location 2 Wall Mount 3 Pole Mount 3 VESA Mount 3 Stand Mount...

Страница 2: ...onics product or system This manual may not be disclosed in any form in whole or in part directly or indirectly to any third party General Information The TempSense Model TS100 measures and captures E...

Страница 3: ...o the camera Camera Location Locate the TempSense TS100 so that people in a queue individually pass through the two cameras shared FOV field of view while facing the cameras Each subject scanned shoul...

Страница 4: ...e the TS100 mounting plate see Figure 4 to a pole using customer supplied worm drive clamps 2 Secure the TS100 to the mounting plate using the four included 6 32 screws VESAMount Attach a VESA video e...

Страница 5: ...100 s IR sensor has a white LED that blinks slowly for 40 seconds then changes to a fast blink NOTE If the IR sensor LED does not flash press the sensor into its socket using a soft cloth then reboot...

Страница 6: ...nce from where the TS100 will scan subjects The calibration object must be the hottest object in the FOV 2 While watching the temperature of the object in the IR video feed left side press the setup b...

Страница 7: ...splays a still image of the IR and standard camera s picture OVER TEMPERATURE flashes on screen notifying the operator of the overtemperature subject The TS100 displays the measured temperature of the...

Отзывы: