3. Use of G end (protection ring) (this machine is shielded on the low voltage side)

When measuring high insulation resistance, a conductor protection ring should be put on

the surface between the two measurement terminals of the test product, and the

conductor protection ring should be connected to the G terminal of the meter with a test

wire to eliminate the leakage current caused by the surface of the test product. The

measurement error ensures the accuracy of the test. Especially when verifying the

instrument, the G terminal should be connected to the G terminal of the resistance box to

ensure normal verification.

G-side usage diagram

VI.Appendix

Resistance dimension: 1000kΩ=1MΩ, 1000MΩ=1GΩ, 1000GΩ=1TΩ;

Current capacity: 1A=1000mA, 1mA=1000uA, 1uA=1000nA;

Содержание HZJY-10K-IA

Страница 1: ...HZJY 10K IA Insulation Resistance Tester User Manual Huazheng Electric Manufacturing Baoding Co Ltd...

Страница 2: ...al and operate and maintain the instrument according to the manual to prolong its service life Just a light press the test will be completed automatically is the operating characteristics of this inst...

Страница 3: ...Contents I Safe Use Items 1 II Features 2 III Technical specifications 3 IV Instrument introduction 4 V Instrument use 5 VI Appendix 7...

Страница 4: ...erious or fatal damage Before use please wear insulated gloves Please do not over AC DC600V measurement circuit Please do not test in flammable places spark could cause an explosion please do not hand...

Страница 5: ...nt and capacitance as well as dielectric discharge test DD step voltage test SV ramp test the instrument automatically boosts Until the set value it is qualified If it is not reached there is a breakd...

Страница 6: ...l value Output voltage monitoring accuracy 5 10V Voltage measurement range AC 30 600V 50HZ 60HZ DC 30 600V Voltage measurement accuracy 2 3dgt Current test range 10mA Current measurement accuracy 5 0...

Страница 7: ...utton 6 High voltage indicator Keys and function description There are 5 buttons in total from left to right The first one is the power on button in the shutdown state Button No Instruction 1 Power on...

Страница 8: ...When charging the power indicator light is on and the battery indicator on the display screen displays from 0 grid to full grid When it is full it displays full grid System help menu description Help...

Страница 9: ...stance values of 15S 60S and 600S and calculates the polarization index after 600 seconds Polarization index calculation method PI R600 R60 DD test This test is suitable for the diagnosis of multilaye...

Страница 10: ...should be connected to the G terminal of the meter with a test wire to eliminate the leakage current caused by the surface of the test product The measurement error ensures the accuracy of the test Es...

Отзывы: