background image

4

A

B

C

I+

I-

Measurement device

O

V+

V-

a

b

c

Testing
product

A

B

C

I+

I-

Measurement device

O

V+

V-

a

b

c

Testing
product

Picture 3, 4, 5 respectively measure low voltage Rac, Rba, Rcb connection

2.

Current options

: open the power switch (“I” opens the switch, “O” closes it) meanwhile all the

current values will be displayed on the screen, then user can select key for the test products’

current preferences, each time the user press a key, the cursor will roll between the current value

of <5mA, 1A, 5A, 10A, 30A.

3.

Testing

:

Time
Data
Current

10A

Pic 5

Содержание HZ-3150A

Страница 1: ...HZ 3150A Transformer DC Resistance Tester USER MANUAL Huazheng Electric Manufacturing Baoding Co Ltd...

Страница 2: ...ual and operate and maintain the instrument according to the manual to prolong its service life Just a light press the test will be completed automatically is the operating characteristics of this ins...

Страница 3: ...s I Introduction 1 II Safety Measure 1 III Configuration 1 IV Technical Data 2 V Presentation Of The System 2 VI Test And Operating 3 VII Caution 5 VIII Guarantee and After Sales Service 5 IX Packing...

Страница 4: ...way to achieve fast transformer measurement II Safety Measure 1 Before using this equipment you must carefully read the manual 2 The equipment operator should have electric general knowledge 3 Avoid...

Страница 5: ...ndensed 7 Working power AC220V 10 50HZ 1HZ V Presentation Of The System Instrument panel see the picture 220V Power switch machine power input AC socket safety and switch Earthing pole indicator use t...

Страница 6: ...vice through test terminal joint firm meanwhile insure a good connection of the earthing wire A Direct measurement connection see picture 2 A B C I I Testing product Measurement device O V V a b c B m...

Страница 7: ...Rac Rba Rcb connection 2 Current options open the power switch I opens the switch O closes it meanwhile all the current values will be displayed on the screen then user can select key for the test pro...

Страница 8: ...e reconnecting wires 3 Before selecting the current refer to the technical instruction range don t exceed or be lower than the use range When over passing the range because the current doesn t reach t...

Страница 9: ...lease timely contact with us and then we will arrange convenient and fast disposal and field service IX Packing List No Item Qty 1 Main engine 1 2 Red test line 1 3 Black test line 1 4 Power line 1 5...