4. Foot control connection
Connect the foot control plug to the foot control connector at the back
of the control unit. (Fig. 6)
Align the marks on the foot control cord and on cowl located on the
back of the unit. Take care to properly align all the pins. Push the plug
gently into the connector..
5. poWer corD connection
Insert the power cord into the inlet at the back of the control unit.
(Fig. 7) Plug the power cord into the power receptacle.
6. hanDpiece corD connection
Align the marks on the control unit and the handpiece cord plug.
Gently push the plug into the connector at the front of the control
unit. (Fig. 8)
7. hanDpiece connection
Confirm there is no moisture in the handpiece connector. If moisture is
present, dry it with cloth or air syringe.
Align the marks on the handpiece with the marks on the handpiece hose
to properly align the electrical contacts. (Fig. 9) Gently push straight to
connect. Do not twist handpiece connector to connect.
Place the handpiece in the handpiece holder on the control unit.
To remove handpiece from the handpiece cord, first remove tip, then
firmly grip the front section of the handpiece in one hand and the rear
section of the handpiece cord plug with the other hand and pull to
separate. Do not twist to separate.
WARNING:
Do not touch the back end of the handpiece (where
electrical connections are made to the cord). It may result in electric shock.
CAUTION:
1. Do not disconnect the cord handpiece when power to the device is switched on and the
footswitch is depressed.
2. Detach tip before removing handpiece.
3. Push handpiece against handpiece cord plug to connect.
4. Grip the plug of the handpiece cord when removing handpiece.
8. start-Up
The following procedures are recommended before using your Symmetry IQ 4000 for the first time:
• Clean and sterilize all appropriate accessories.
• After installation operate the device at the lowest power setting while applying pressure to the foot
control. This procedure will confirm adequate water flow and to rinse the internal water circuits.
• Disconnect the handpiece and proceed with recommended cleaning and sterilization procedures.
(See pages 12-13, Cleaning & Sterilization)