INSTALLATION AND ASSEMBLY
1. Unpacking the Device
Upon receipt please check device for any damage caused in transit. If necessary, contact your dealer.
2. general recommenDations
Connect the Symmetry IQ 4000 control unit to electric supply in accordance with current standards.
The water supply system should be of adequate quality for the practice of dental treatment. Additionally, the
city water supply should be equipped with a cut-off system.
3. Water system setUp
Use of City Water
Remove the cover from the water connector (Fig. 1)
Connect the filter side of the water tube by inserting deep into the
water connector on the Symmetry IQ control unit. (Fig. 2)
Connect the provided water tube to the water outlet on the dental unit.
To remove the water tube from the control unit, push the white ring
(the quick connector release ring) and gently remove the tube. (Fig. 3)
When the water tube is not connected, place the cover on the water
tube connector on the control unit.
CAUTION:
Be sure to insert the water tube deep into the
connector on the control unit. If water has not been used recently
at the water outlet of the dental unit, discolored water may come
out. Wait until the water coming out of the dental unit is clean
before connecting to the control unit.
Use of Irrigation Bottle
Remove the dust cover from the base unit. (Fig. 4)
Remove the cap of the irrigation bottle and fill solution to the desired
level.
Close the cap of the irrigation bottle, check the air hole to confirm
it is clean and insert the bottle into the connector located on the unit.
(Fig. 5) When attached properly, the irrigation bottle will click into place.
To remove the irrigation bottle, pull the bottle straight up.
Gradation lines are printed on both sides of the irrigation bottle and
can be read accurately from the fill position or when mounted on the
control unit.
Mount the dust cover on the base unit when irrigation bottle is not in
use.
NOTICE:
Please make sure the connector and surrounding area
is completely clean before installing the irrigation bottle. Wipe the
irrigation bottle and connector area thoroughly prior to inserting
the irrigation bottle.