44
Cleaning and disinfecting the body and body cord
Clean the surface of the body and body cord using a cleaning wipe with proven ef
fi
cacy (Enzol 2%).
Disinfect the surface of the body and body cord using 70% isopropyl alcohol or other EPA recognized surface disinfectant, following carefully the instructions
provided by the disinfection solution manufacturer.
The use of scouring powder or an abrasive sponge will damage its surface.
Do not submerge the body in a cleaning solution bath. It would be likely to be damaged.
Make sure the handpiece connection of the cord is dry.