28.25.14−001−89695785−2020 PS
6 из 12
Troubleshooting
T a b l e 5 - Troubleshooting
Problem
Possible Cause
Corrective Action
The Power indicator is not lit
The fuse is faulty
Replace the fuse
The UV indicators are blue (the
UV lamps are on), the fan is not
working
The fan failed
Replace the fan
The fan is working, the UV lamp
indicator(s) is(are) not lit
There is no contact between the
UV lamp and the lamp holder
Reinstall the UV lamp in the
holder
The UV lamp inoperative
Replace the UV lamp
The ECG (electronic ballast) is
faulty
Replace the ECG (electronic
ballast)
10
Shipping and storage
1.
The Recirculator is shipped by all types of transport in accordance with the rules applicable to this
transport.
2.
The Recirculator keeps workability under the vibration loads with a frequency of up to 20 Hz at 0.5
g acceleration.
3.
The Recirculator is shipped in individual or multipack containers. The shipping container must have
an inscription: "Watch the glass."
4.
The minimum and maximum permissible ambient temperature and humidity for transportation and
storage would be specified additionally in the shipping documents if necessary.
5.
Climatic terms for the Recirculator transportation must correspond to the storage conditions 3 (G3)
as per GOST 15150-69 at temperatures from -40 ° C to + 60 ° C, relative humidity of 80% at + 25
°C.
6.
The height of the storage pile should not exceed 2.7 m during transportation.
7.
Container
with
the
Recirculator
must
be
stored
indoors
at
temperatures
from -40 ° C to + 60 ° C, relative humidity of 90% at + 25 °C. A storage room must be protected
from precipitations, direct sunlight and aggressive environments.
8.
The height of the storage pile should not exceed 1.7 m at the Recirculator boxes stowing. Boxes
with the Recirculator should be placed on pallets, racks or decking at 0.12 m from the floor and
external walls as minimum.
It is forbidden to place other types of cargo on the Recirculator cardboard box.
9.
In order to avoid the mechanical damaging, follow the loading/unloading precautions.
Loading and unloading of the equipment must be executed according to the GOST 12.3.009-76
smoothly without jerking and striking.
Dropping from vehicles is not permitted.
10.
Packaged Recirculators must be stored on pallets at least 1 m away from the heating devices.
11
Manufacture’s warranty
1.
The manufacturer warrants The Recirculator quality in accordance with the SPEC. 28.25.14-001-
89695785-2020 while it proper operated, stored and transported.
2.
Warranty period is 12 months from the date of manufacture.
It is allowed to set warranty periods in vendor contracts.
3.
The Recirculator or the failed parts are repaired by the Manufacturer during the warranty period.
4.
The Manufacturer’s (Supplier) warranty becomes not valid in the case of:
-
non-observance of operating rules by owners;
-
lamp failure;
-
careless storage and transportation;
-
use of the Recirculator for other purposes which is not intended for;
-
in the presence of mechanical damage to the recirculator caused by external factors;
-
in case of malfunctions that have arisen as a result of exceeding the parameters specified in the
passport, as well as a result of a mismatch between the parameters of the electrical network and the
rated voltage;
Содержание HTL-Advansys-15.2U
Страница 12: ...28 25 14 001 89695785 2020 PS 12 из 12 ...