Operating procedure  

The Recirculator operating time should be recorded in  the Readings and Monitoring  Log. Results  of 
visual  observations  of  the  germicidal  lamp  surface  cleanliness  should  be  recorded  periodically  to 
determine the optimal period for the Recirculator preventive maintenance. 

 
 

Operation of the device in the closed-type air recirculator-irradiator mode:  

1) Place the recirculator on a hard surface using a stand in an open space of at least 1 m²; 
2)      If it is necessary to install on a wall, it is necessary to dismantle the stand by first unscrewing the 
bolts of the stand clamps; 
3)      Screw the hangers (brackets) supplied with the recirculator into the holes to fix them. 
4)      Connect the cable to the power connector; 
5)      Turn on the power button of the recirculator. 
 

 

Operation of the device in the open-type air recirculator-irradiator mode: 

1)

 

Turn off the Recirculator with the power button 1; 

2)

 

Install the device with its face towards the surfaces to be irradiated; 

3)

 

Open the reactor chamber door; 

4)

 

 

Step out of direct exposure to UV radiation (behind the unit) as shown in figure 1; 

5)

 

Put the safety glasses on; 

6)

 

Turn on the device in the irradiation mode by pressing the power button 1; 

7)

 

Leave the room for the entire period of the irradiator operation. The recommended operating 

time of the Recirculator in open mode is 15-90 minutes for a 20 m² room. 

 

 

 

Figure 1 

 

 

WARNING 

In any case Do not look at the emitters in order to avoid the eye cornea  damage! 

 
 
9.

 

Maintenance  

The maintenance of the Recirculator should be carried out by the Manufacturer or by a certified 
service facility only.  
 

 

WARNING 

DIY troubleshooting, repair and adjustment are not allowed

.  

 

 
 
 

Содержание HTL-Advansys-15.2U

Страница 1: ...nts of GOST 2 601 HTL ADVANSYS LLC OKPD2 Russian Product Classification Code 28 25 14 110 EAC Bactericidal Air Recirculator Irradiator Closed opened type HTL Advansys 15 2U Technical Passport Operators Manual 28 25 14 001 89695785 2020 PS ...

Страница 2: ...parameters of the Recirculator are carried out constantly 2 Operational and technical specifications 2 1 The main specifications of the Recirculator are shown in the Table 1 Table 1 Basic technical data Parameter Value HTL Advansys 15 2U HTL Advansys 15 2U with a stand Capacity m3 per hour 70 Total power W 45 Power supply Voltage V 230 10 Power supply Frequency Hz 50 Disinfection efficiency not le...

Страница 3: ...G1 into the reactor chamber where it is disinfected by receiving the necessary dose of UV radiation when moving along the bactericidal lamp 6 Pre start procedure 1 Remove the Recirculator from the shipping container Let it stay at room temperature for 2 3 hours before switching it ON if the Recirculator has been transported at sub zero temperatures 2 Install the Recirculator on an open surface or ...

Страница 4: ...2 The Manufacturer guarantees the Recirculator safety in accordance with the requirements of Spec 28 25 14 001 89695785 2020 and TR CU 020 2011 The Customs Union Technical regulations for Electromagnetic compatibility of technical facilities Low voltage equipment safety 3 The Recirculator is not intended for use by persons with disabilities reduced physical sensory or mental capabilities or person...

Страница 5: ... power connector 5 Turn on the power button of the recirculator Operation of the device in the open type air recirculator irradiator mode 1 Turn off the Recirculator with the power button 1 2 Install the device with its face towards the surfaces to be irradiated 3 Open the reactor chamber door 4 Step out of direct exposure to UV radiation behind the unit as shown in figure 1 5 Put the safety glass...

Страница 6: ...ity of 90 at 25 C A storage room must be protected from precipitations direct sunlight and aggressive environments 8 The height of the storage pile should not exceed 1 7 m at the Recirculator boxes stowing Boxes with the Recirculator should be placed on pallets racks or decking at 0 12 m from the floor and external walls as minimum It is forbidden to place other types of cargo on the Recirculator ...

Страница 7: ...program and methodology or manufacturer s recommendations 4 It is recommended to utilize The Recirculator as consumer waste to the recyclable materials and resources The Recirculator must be disposed in compliance with the requirements listed in paragraph 13 5 of this Section 5 If consumer decides to utilize The Recirculator and its components he must follow the requirements of waste management an...

Страница 8: ...ct from the power supply cleaning from pollutions etc 9 At the sorting stage identified and certified elements are preliminarily divided according to the specified criteria into qualitatively different components in the form of recyclable materials for example metals plastics etc The body and other metal parts of The Recirculator are recyclable as secondary metal resources ...

Страница 9: ...n Date of manufacture Person responsible for the acceptance position signature surname date Stamp here Acc to the Agreement Agreement from 202__ if applicable delivered on the base of designated document name of the document If any additional information is required please contact us Contacts Manufacture HTL Advansys LLC Manufacture address Russia Telephone 371 28 300 003 7 929 685 66 41 e mail HT...

Страница 10: ...ation are entered into the Table 7 Table 7 Records of The Recirculator failure during the operation Date of failure The nature external manifestation of the fault Operation time period of operation Adequate corrective actions Position initials surname and signature of the responsible person Notices 17 Records of The Recirculator Operating Time before repair 1 Records of The Recirculator Operating ...

Страница 11: ...ompletion of the repair 20 Special notices The passport must not contain writing in pencil dilutable ink or erasures The wrong entry should be carefully crossed out and the new one should be written next to it All the new entries must be certified by the responsible person After the signature the surname and initials of the responsible person must follow it is allowed to put the performer s person...

Страница 12: ...28 25 14 001 89695785 2020 PS 12 из 12 ...

Отзывы: