DIGITAL MULTIMETER USERS MANUAL 

 

1.

 

GENERAL DESCRIPTION 

This  31/2  Digit  Highly  Stable  multimeter  uses  two  AA  (1.5V)  battery.  It  has  a  LCD  with 
22mm  digit  display,  which  makes  the  reading  clearer  and  the  operation  more  convenient.  It 
can  test  DCV,  ACV,  DCA,  ACA,  resistance,  NCV,  temperature,  transistor,  diode,  and 
continuity  test.  This  meter  also  has  Max  Min  Hold  function  with  unit  symbol  display,  can 
work  in  both  auto  and  manual  range,  and  provides  auto  power  off,  backlight  and  warning 
functions. To assure high accuracy and resolution, it adopts an 8-bit microprocessor and a dual 
integral A/D convertor IC .It is an ideal tool for labs, factories and radio-technology.

 

2.

 

SAFETY PRECAUTIONS 

The  instrument  is  designed  according  to  IEC1010  standard  (safety  standard  issued  by 
International  Electro  technical  Committee).  Please  read  the  following  carefully  before 
operation. 
2.1 Check the connection and    insulation of test leads to avoid electric shock. 
2.2  To  avoid  electric  shock  and  damage  to  the  meter,  do  not  input  voltage  higher  than  DC 
1000V or AC 750V during measurement. 
2.3 Use caution when working with voltages above DC 60V or AC 40V. 
2.4 Select correct function and range to avoid shock and fault operation. 
2.5 Please move the test leads away from test points when switching the Range. 
2.6 Please don’t input voltage in current terminal. 
2.7 Introduction for safety symbols: 
  “

”  exists  high  voltage,  “ ”  GND,  “ ”  dual  insulation,  “ ”  Operator  must  refer  to 

manual, “

” Low battery indication. 

3.

 

FEATURES 

3.1 General Characteristics 

3.1.1 Display: LCD; 
3.1.2 Max display: 2,000 (3 1/2 digits, auto polarity, and unit symbol display); 
3.1.3 Measurement method: Analog to digital converter (in micro processor ADC+MCU); 
3.1.4 Sampling rate: approx.3 times/sec. 
3.1.5 Over-range display: “OL” displayed. 
3.1.6 Low battery indicator: “

” 

3.1.7 Working environment: (0

40)℃, relative humidity: <80%; 

3.1.8 Store condition: (-10

50)℃, relative humidity: <80% 

3.1.9 Battery: 2 X 1.5V battery (AAA); 
3.1.10 Dimension: 170×86×35mm (length*width*height); 
3.1.11 Weight: approx. 290g (including battery); 
3.1.12 Accessories: test leads, user manual, temperature probe, gift box and batteries. 

3.2 Technical Features 

3.2.1  Accuracy:  ±  (a%  ×  reading  data  +  digits).  To  assure  accuracy,  the  environment 
temperature should be (23±5) 

, relative humidity should <75%. 

One year accuracy guarantee since production date. 
3.2.2 DC Voltage

DCV

 

Range 

Accuracy 

Resolution 

200mV 

±(0.5%+4) 

0.1mV 

2V 

1mV 

20V 

10mV 

200V 

100mV 

1000V 

±(1.0%+4) 

1V 

Input impedance: at 200mv range >40MΩ, at other ranges is 10MΩ. 

Overload protection: 1000V DC or 750V AC peak value 

3.2.3 AC Voltage

ACV

 

Range 

Accuracy 

Resolution 

2V 

±(0.8%+6) 

1mV 

20V 

10mV 

200V 

100mV 

750V 

±(1.0%+6) 

1V 

Input impedance: 10MΩ. 

Overload protection: 1000V DC or 750V AC rms 

Frequency response: at 750V range: 40~1000Hz, at other ranges: 40~2000Hz 

Displaying: True RMS response (calibration based on sine wave RMS)

 

3.2.4 DC Current

DCA

 

Range 

Accuracy 

Resolution 

200uA 

±(1.0%+5) 

0.1μA 

2000uA 

1μA 

20mA 

10μA 

200mA 

100μA 

2A 

1mA 

20A 

±(2.0%+5) 

10mA 

Maximum voltage drop: 200mV for all ranges. 
Maximum input current: 20A (for 15 seconds). 
Over load protection: 0.2A/250V fuse and 13A/250V fuse. 
3.2.5 AC Current

ACA

 

Range 

Accuracy 

Resolution 

200uA 

±(1.5%+5) 

0.1μA 

2000uA 

1μA 

20mA 

10μA 

200mA 

100μA 

2A 

1mA 

20A 

±(2.0%+10) 

10mA 

Maximum voltage drop: 200mV for all ranges. 
Maximum input current: 20A (for 15 seconds). 
Over load protection: 0.2A/250V fuse and 13A/250V fuse. 
Frequency response: 40~100Hz under 20A range, 40

400Hz at other ranges. 

3.2.6 Resistance

Ω

 

Range 

Accuracy 

Resolution 

200Ω 

±(0.8%+5) 

0.1Ω 

2kΩ 

±(0.8%+1) 

1Ω 

20kΩ 

10Ω 

200kΩ 

100Ω 

2MΩ 

1kΩ 

20MΩ 

±(1.2%+5) 

10kΩ 

Open circuit voltage: >500mV 
Overload protection: 250V DC/AC peak value. 
3.2.7 Transistor triode (hFE) 

Measurement 

Range 

Test Conditions 

NPN or PNP 

0

1000 

Base current is approx 

15uA, Vce is about 1.2V 

3.2.8 Diode and Continuity Test 

Range 

Description 

Test Conditions 

 

Diode forward voltage drop 

Forward DC current is 

approx 0.8mA, reverse 

voltage is approx 2.2V. 

When the resistance under test is 

less than 50Ω, buzzer sounds 

continuously. 

Open circuit voltage: 2.2V 

Overload protection: 250V DC/AC peak value 
WARNING: Do not apply any voltage in this range. 
3.2.9 Temperature (

/

Range 

Accuracy 

Resolution 

-40℃~1000℃ 

400℃ ±(1.0%+5) 

≥400℃ ±(1.5%+15) 

1℃ 

0F~1832℉ 

750℉ ±(1.0%+5) 

≥750℉ ±(1.5%+15) 

℉ 

Thermocouple: K type 
WARNING: do not input voltage at this range! 
 

OPERATION INSTRUCTION 

4.1 Panel Description

 (See picture) 

. NCV detector area and indicator light. 

.LCD: Display the data and unit symbol. 

 

 

 

 

 

 

 

 

Number 

Feature 

Indication 

APO 

Auto power off symbol. 

 

Low battery indication. Warning: To avoid error readings, 

which could lead to possible electric shock or personal 

injury, please replace the battery in time. 

Indicates negative readings. 

AC 

AC voltage or current measurement. 

DC 

DC voltage or current measurement. 

AUTO 

Auto range mode. 

 

Diode test mode 

 

Continuity beeper is on. 

HOLD 

Display Hold is active. 

10 

MAX MIN 

 

Max&Minimum value recording mode. 

Null 

11 

hFE 

℃/℉ 

MΩ, kΩ, Ω 

mV, V 

uA, mA, A 

NCV 

hFE (Triode magnification measurement) 

Degrees Celsius, Degrees Fahrenheit 

Megohm, Kilohm, Ohm 

Millivolts, Volts 

Microamp, Milliamp, Amperes (A) 

Non-contact voltage detector. 

Function Button 

-1 

  LIGHT  key:  press  “ ”  light  key  to  turn on  the  backlight,  the  backlight  can  last  15 

seconds. During 15 seconds, press “ ” light key again will turn off the backlight.   

-2 MAX / MIN: Recording of minimum and maximum values 

1) Press MAX/MIN into MAX mode, which will store the maximum value of measurements; 
Press it again to start MIN mode, which will store the minimum value of measurements. 
2)  Under  MAX/MIN  mode,  it  will  start  manual  range  automatically.  Under  this  mode, 
functions like HOLD, RANGE, and DC/AC will not be available. 
3) Under MAX/MIN mode, it will store maximum or minimum values automatically.   
4) Under MAX/MIN mode, auto power off can’t be used. 
5) Press MAX/MIN for 2 seconds it will exit MAX/MIN mode. 

-3 HOLD key: press HOLD key to enter HOLD mode, the current value will be hold, and 

symbol HOLD will be displayed. Press HOLD again can exit the HOLD mode. 

-4  RANGE  key:  for  choosing  auto  or  manual  range.  The  meter  defaults  to  auto  range  in 

measurement function and “AUTO” symbol is displayed. To Enter Manual Range mode Press 
RANGE  key.  Each  Press  of  RANGE  key  Increments  the  range,  when  the  highest  range  is 
reached ,the meter wraps to the lowest range .To exit the manual range mode press and hold 
the Range key for more than 2 seconds.   

-5 SELECT key (yellow):   

-5-1 At 

  range, press yellow key can choose DC or AC measurement. At 

  range, 

press yellow key can choose diode or continuity test. At temperature range, press yellow key 
can choose 

/

-5-2 When there is no measurement for 15 minutes, the meter will auto power off and enter 

sleep  mode.  In  one  minute  before  Sleep  mode,  the  buzzer  will  beep  5  times  to  remind  user. 
Press any button or turn the rotary switch will exit the Sleep mode. Press yellow key to Power 
on the meter from the Sleep mode or Press and hold the yellow button when the meter is ON , 
it will cancel auto power off function. 

.Rotary switch: used to change the range and choose functions. 