disturbance and might cause the wrong results.   

4.6 DIODE AND CONTINUITY TEST 

4.6.1 Set the rotary switch to “

” range. The 

original state is diode measurement mode. (Refer to 
picture 6)

 

4.6.2 Forward measurement: Connect red test lead to 
the positive polarity and the black test lead to the 
cathode polarity of the diode. LCD will display the 
approx. value of forward voltage drop.                 (picture 6) 
4.6.3 Backward measurement: connect red test lead to the cathode 
polarity and the black test lead to positive polarity of the diode. LCD 
will display “OL”. 
4.6.4 The complete diode testing includes forward and backward 
measurement. If the result doesn’t meet the descriptions above, it means 
the diode is broken. 
4.6.5 Press SELECT (yellow) key to select the continuity measurement 
mode. 
4.6.6 Insert the black test lead into “COM” terminal, and the red one 
into “V/

” terminal. 

4.6.7 Connect test leads to two points of tested circuit, if the resistance 
is less than (50±10) 

, the buzzer sounds. 

Note:

 DO NOT input any voltage value at this range! 

4.

7 RESISTANCE MEASUREMENT 

4.7.1 Turn the function swift to “

” range. Press “SELECT” 

to turn to “

” measurement mode. 

4.7.2 Insert the black lest lead into “COM” terminal and the red one 
into “V/

” terminal. 

4.7.3 Connect the test leads to the measured resistance to get the 
measurement value. (Refer to picture 6)

 

Note: 

1. Before measuring in line resistor, make sure that the power is off and 
all capacitors have discharged completely.

 

2. Do not apply any voltage at resistance range. 
3. When input terminal is in open circuit, LCD will display “OL”. 
4. When the test leads are shorted, there would be about 1

 stray 

resistance. 

 

4.8 CAPACITANCE MEASUREMENT 

4.8.1 Turn the function swift to “

” range. Press “SELECT” 

to turn to “

” measurement mode. 

4.8.2 Insert the black lest lead in “COM” terminal and the red one in 
“V/

” terminal.

 

4.8.3 If the LCD doesn’t display “0”, press “REL” to clear the reading; 
4.8.4 Connect the capacitor to “COM” and “V/

” terminal. (Note: the 

red test leads is for positive pole +). LCD displays capacitance value. 

Note: 

1. Don’t input voltage or current to the “V/

/” terminal when 

measuring the capacitance or the capacitor still in the “Cx” terminal. 
2. In order to assure the accuracy, please press “REL” to clear the 
reading before testing. 
3. There is only the auto range mode under the capacitance range. 
4. The capacitor must be completely discharged before testing. 
5. The reading of 200uF range will take more than 15 seconds to be 
stable. 

4.9 FREQUENCY MEASUREMENT 

4.9.1 Turn the function swift to Hz range. Insert the black lest lead in 
“COM” terminal and the red one in “V/

” terminal. 

4.9.2 Connect the test leads to the measured circuit 
and get the measurement value. (Refer to picture 7)

 

Note: 

1. In noisy environment, it's better to use shield 
cable to measure a low signal. 
2. When measuring high voltage circuit, do not touch the high voltage 
circuit; 
3. Don’t input voltage higher than 250V TRMS or it may damage the 
meter. 

4.10 TEMPERATURE TEST 

4.1.11 Set the rotary switch to “

/

” range. 

4.1.12 Insert the black plug of the thermocouple 
sensor into “COM” and red one into “V/

” jack, 

and put the working terminal (temperature 
measuring end) of thermocouple on the surface or 
inside the object to be tested. Then you can read 
temperature from the screen, and the data is in 
Centigrade. (Refer to picture 8) 

Note

1

. The meter should be far away high temperature. Use the 

thermocouple probe within the specified temperature range. 
2. Don’t change the temperature probe randomly, or the accuracy will 

not be guaranteed.   
3. Don’t input voltage at temperature range. 

4.11 AUTO POWER OFF 

4.11.1 The meter will auto power off if there is not any operation in 15 
minutes and come into dormant mode. The beeper will sound five times 
within 1 minute before power-off. 
To restart the meter, please press any function key or turn to the 
function swift. 
4.11.2 Press “SELECT” key and turn on the meter or press “SELECT” 
when the meter is in dormant mode to wake up the meter and the 
“AUTO POWER OFF” function will be cancelled 

5. MAINTENANCE 

The meter is a precise instrument. Random changes to the circuit should 

be avoided. 

Note: 

1. When the LCD displays “

”, please exchange the battery in time.

 

2. Before replacing battery or fuse, please remove the test leads from 

the measuring point and turn off the power. 

3. Keep the meter away from water, dust and shock. 

4. DO NOT use the meter under high temperature, high humidity, 

combustible, explosive and strong magnetic environments. 

5. Clean the case of the meter regularly with cleaner. DO NOT use 

corrosive solvents.   

6. TROUBLE SHOOTING 

If the meter does not work properly, please check the meter as 
following steps: 
(If the problems still cannot be solved, please refer to repairing center 
or contact the local dealers.) 

Fault Solution 

No reading on LCD 

„

 

Turn on the power 

„

 

Replace battery 

„

 

Release the HOLD key 

 signal appears 

„

 

Replace battery 

No current or temperature input 

„

 

Replace fuse 

Big error Value 

„

 

Replace battery 