830L+    Digital Multimeter Operation Manual

 

 

It’s very helpful that reading the operation manual to familiar with how to operate and maintenance the device 

before operation.   

 

1.

 

Appearance Introduction 

1-1.

 

Display 

3

½

 LCD with 20mm high figure display. 

 
1-2.

 

Function Keys 
SELECT: 

select 

  or 

  function 

under 

  status. The device goes to sleep when stopping the operation for 15 minutes, 

and re-powers on when press the function key or toggle the function switch.

 

 
HOLD/

:   

 

Data hold: press this key for a short time, the current measuring value of the device will be hold on the LCD and “H” symbol 
will be displayed. Press it again can exit the HOLD mode. 

 

Backlight: press this key for more than 2 seconds to turn on the LCD backlight. The LCD backlight will be turned off 
automatically after 15 seconds, or turned off by press this key again for more than 2 seconds. 

 
1-3.

 

Range rotary switch               
This device is adapting rotary type switch, it combines function selection, range selection and power switch into 
one role. To extend the battery life, the rotary switch should be turned to “OFF” position if the device is not be 
used. 

 
1-4.

 

Input Terminal 
“VΩmA”: Input terminal for Voltage, Resistance and less than 200mA current. 
“COM”: Input terminal for common GND. 
“10A”: Input terminal for more than 200mA current. 
“NPN/PNP”: Input terminal for triode measurement. 

 
1-5.

 

The top of the LCD is non-contact voltage (NCV) detector area. 

 
1-6.

 

Holster and battery door. 

 

2.

 

General Features 

Display: 3

½

 LCD, Max. display 1999. 

Polarity: Automatic polarity display. 
Over-range display: Display “1”. 
Working environment: (0~40)

, relative humidity <75%.   

Storage environment: (-15~50) 

Battery: 2pcs 1.5V battery (“AAA” 7# battery). 
Battery life: About 200 hours for alkaline battery, and 100 hours for carbon battery. 
Dimension&Weight: 140mm(L) x 72mm(W) x 37mm(H), N.W: 195g 
Low battery indication: LCD displays ”

 

 

3.

 

Technical Features 

Environment condition: Temperature: 23

°

C±5

°

C, Relative humidity: <75%. Accuracy: ± (a% × reading data + digits), one 

year guarantee since production date. 
 
3-1. DCV 

            Range 

            Resolution 

              Accuracy 

200mV 

100uV 

±(0.5% r 4) 

2000mV 

1mV 

20V 

10mV 

200V 

100mV 

600V 

1V 

Input impedance: 200mV range >40MΩ, other ranges is 10MΩ. 
Overload protection: 600V DC/AC RMS. 

3-2. DCA 

              Range 

Resolution 

Accuracy 

20uA 

10nA 

±(1.0% r 5) 

2mA 

10uA 

200mA 

100uA 

±(2.0% r 5) 

10A 

10mA 

Overload protection: 0.2A/250V fuse, 10A/250V fuse. 

 

3-3. ACV 

              Range 

Resolution 

Accuracy 

200V 

100mV 

±(1.0% r 6) 

600V 

1V 

Frequency response: 40-400Hz. 
Overload protection: 600V DC RMS or AC RMS. 
Displaying: True RMS response (calibration based on sine wave RMS).   
 
3-4. Resistance 

              Range 

Resolution 

Accuracy 

200Ω

 

0.1Ω

 

±(0.8% r 5) 

2kΩ

 

 

±(0.8% r 1) 

20kΩ

 

10Ω

 

2MΩ

 

1kΩ

 

200MΩ

 

100kΩ

 

±(1.2% r 5) 

Open circuit voltage: 

500mV 

Overload protection: 250V DC/AC peak value. 
 
3-5. Transistor (hFE) 

Measurement 

Range 

Test conditions 

hFE NPN or PNP 

0

1000 

Base current is approx 15μA, Vce is 
about 1.2V 

 

3-6. Diode and Continuity Test 

Measurement 

Range 

Test conditions 

 

Diode forward voltage drop 

Forward DC current: 0.8mA, open circuit 
voltage: 2.2V.

 

When the resistance under 
test is less than 50Ω, buzzer 
sounds continuously. 

Open circuit voltage: 2V 

Overload protection: 250V DC/AC peak value. 
WARNING: Do not input any voltage at this range! 

 

4.

 

OPERATION 

4-1. DCV measurement 

A.

 

Insert the red test lead into “VΩmA” terminal, and the black one into “COM” terminal; 

B.

 

Turn the rotary switch to DCV position and select the proper range, if users not sure about the range, then 

select the highest range; 

C.

 

Connect the red and the black test leads to the tested circuit. 

 

4-2. DCA measurement 

A.

 

Insert the black test lead into “COM” terminal and the red one into “VΩmA” terminal (less than 200mA) or 

into ”10A” terminal (more than 200mA); 

B.

 

Turn the rotary switch to the proper DCA range position; 

C.

 

Connect the red and the black test leads to the tested circuit. 

 

4-3. ACV measurement (TRMS) 

A.

 

Insert the red test lead into “VΩmA” terminal, and the black one into “COM” terminal; 