Emmisivity Values

Substance

Asphalt

Concrete

Cement

Sand

Earth

Water

Ice

Snow

Glass

Ceramic

Marble

Plaster

Mortar

Brick

Thermal Emissivity

0.90 to 0.98

0.94

0.96

0.90

0.92 to 0.96

0.92 to 0.96

0.96 to 0.98

0.83

0.90 to 0.95

0.90 to 0.94

0.94

0.80 to 0.90

0.89 to 0.91

0.93 to 0.96

Substance

Cloth (black)

Human skin

Lather

Charcoal (powder)

Lacquer

Lacquer (matt)

Rubber (black)

Plastic

Timber

Paper

Chromium oxides

Copper oxides

Iron oxides

Textiles

Thermal Emissivity

0.98

0.98

0.75 to 0.80

0.96

0.80 to 0.95

0.97

0.94

0.85 to 0.95

0.90

0.70 to 0.94

0.81.

0.78

0.78 to 0.82

0.90

3

INTRODUCTION

Thank you for purchase of the 8858/8859 Series IR Thermometer. The 
8858/8859 Series is capable of non-contact (infrared) temperature 
measurements at the touch of a button. The built-in laser pointer 
increases target accuracy while the backlight LCD and handy push-
buttons combine for convenient, ergonomic operation.

The Non-contact Infrared Thermometers can be used to measure the 
temperature of objects’ surface that is improper to be measured by 
traditional (contact) thermometer (such as moving object, the surface 
with electricity current or the objects which are uneasy to be touched.)

Proper use and care of this meter will provide years of reliable service.

FEATURES:

Precise non-contact measurements

High distance to target ratio measures smaller surface areas at
greater distances 

Widest temperature range

Unique flat surface, modern housing design

Built-in laser pointer

Automatic Data Hold

°C

/°F switch

Emissivity Digitally adjustable from 0.10 to 1.0

MAX, MIN, DIF,   AVG temperature displays

Backlight LCD display

Built-in laser pointer

Automatic selection range and Display Resolution 0.1°C(0.1°F)

Trigger lock

Set high and low alarms

Most (90% of typical applications) organic materials and painted 

or oxidized surfaces have an emissivity of 0.95 (pre-set in the 

unit). Inaccurate readings will result from measuring shiny or 

polished metal surfaces. To compensate, cove the surface to be 

measured with masking tape or flat black paint. Allow time for 

the tape to reach the same temperature as the material 

underneath it. Measure the temperature of the tape or painted 

surface.

Содержание EQ-8859

Страница 1: ...Non Contact Infrared Thermometers 8859 series Operation Manual ...

Страница 2: ...ONS 08 7 MEASUREMENT OPERATION 10 8 BATTERY REPLACEMENT 11 9 NOTES 12 10 MAINTENANCE CLEARING 15 03 04 2 15 10 MAINTENANCE CLEARING Repairs or serving aren t covered in this manual should only be performed by qualified personnel Periodically wipe the case with a dry cloth Don t use abrasives or solvents on this instrument When serving use only specified replacement parts ...

Страница 3: ...ture of objects surface that is improper to be measured by traditional contact thermometer such as moving object the surface with electricity current or the objects which are uneasy to be touched Proper use and care of this meter will provide years of reliable service FEATURES Precise non contact measurements High distance to target ratio measures smaller surface areas at greater distances Widest ...

Страница 4: ...ot aim the thermometer outside the area of interest then scan across with an up and down motion until you locate hot spot Reminders Not recommended for use in measuring shiny or polished metal surfaces stainless steel aluminum etc See Emissivity The unit cannot measure through transparent surfaces such as glass It will measure the surface temperature of the glass instead Steam dust smoke etc Can p...

Страница 5: ...e The smaller the target the closer you should be to it When accuracy is critical make sure the target is at least twice as large as the spot size Fig 1 D S 16mm Laser Beam inch 1 2 60 0 72 36 0 24 12 6 0 300 18 900 30 1500 mm Sensor Beam Distance D to Spot Size S D S 50 1 3 SPECIFICATIONS Model Temperature range D S ST 8859 50 to 1600 C 58 to 2912 F 50 1 Display resolution 0 1 C 0 1 F Accuracy fo...

Страница 6: ...it When accuracy is critical make sure the target is at least twice as large as the spot size The meter will automatically power down after approximately 7 seconds after the trigger is released Unless the unit is locked on Note Measurement considerations Holding the meter by its handle point the IR Sensor toward the object whose temperature is to be measured The meter automatically compensates for...

Страница 7: ... buttons to adjust the desired values 7 MEASUREMENT OPERATION Hold the meter by its Handle Grip and point it toward the surface to be measured Pull and hold the Trigger to turn the meter on and begin testing The display will light if the battery is good Replace the battery if the display does not light While measuring the SCAN display icon will appear ins the upper left hand corner of the LCD Rele...

Страница 8: ...mometer measures Maximum MAX Minimum MIN Differential DIF and Average AVG Temp Each time you take a reading This data is stored and can be recalled with the MODE button until a new measurement is taken When the trigger is pulled again the unit will begin measuring in the last mode selected Pressing the MODE button also allows you to access the High MODE Alarm HAL Low Alarm HAL Emissivity EMS Each ...

Отзывы: