PVCHECK
10
5.1.2. Measuring unit
This section allows setting the measuring units of some parameters included in the
database (DB) for the management of PV modules (see § 5.6) when measuring IVCK.
1.
Position the cursor onto “
Measuring Unit
” by using the
arrow keys
and confirm with
ENTER
.
2. The display shows the screen which allows setting the
measuring units of the following parameters:
Alpha
possible selection: “%/°C” and “mA/°C”
Beta
possible selection: “%/°C” and “mV/°C”
Gamma
expressed in “%/°C”
Voc and Isc tolerance
ex
pressed in “%”
3. Set the desired measuring units by means of the arrow
keys
4. Press the
SAVE
key to save the settings made; the
message “Data saved” will be displayed for a few seconds.
5. Press the
ESC/MENU
key to exit without saving and go
back to the previous screen.
5.1.3. Date and time
1.
Position the cursor onto “
Date and Time
” by using the arrow
keys
and confirm with
ENTER
.
2. The display shows the screen which allows setting the
system’s date/time both in the
European (EU)
or in the
USA
(US)
format.
3. Set the desired measuring units by means of the arrow
keys
4. Press the
SAVE
key to save the settings made; the
message “Data saved” will be displayed for a few seconds.
Press the
ESC/MENU
key to exit without saving and go
back to the previous screen.
5.1.4. Remote unit/Pyranometer
This section allows selecting the type of remote unit to be used (if available) and setting the
values of typical parameters (Sensitivity and Alpha) of the reference irradiance cell supplied
with the instrument.
The values of these parameters, which are printed on the back
label of the cell, depend on the type of PV module being tested.
1.
Position the cursor onto
Remote Unit-Pyranometer
by
using the arrow keys
and confirm with
ENTER
.
2.
The display shows the screen which allows selecting the use
of the remote unit for EFF or IVCK tests. Possible options
are:
YES (use of SOLAR-02)
NO (no use of SOLAR-02) In case the remote unit
SOLAR-02 is not used for
IVCK tests
, it is necessary to
set the values of Sensitivity (Sens.) and Alpha parameter
of the supplied irradiance cell on the instrument.
3.
Set the desired values by means of the keys
)
4.
Press the
SAVE
key to save the settings made; the message
“Data saved” will be displayed for a few seconds. Press the
ESC/MENU
key to exit without saving and go back to the
previous screen