HT64
EN - 22
4.4.6. Capacitance
measurement
CAUTION
Before carrying out capacitance measurements on circuits or capacitors, cut
off power supply from the circuit being tested and let all capacitance in it be
discharged. When connecting the multimeter and the capacitance to be
measured, respect the correct polarity (when required).
Fig. 27: Use of the instrument for Capacitance measurement
1. Select
position
CAP
2. Press
the
F2 (MODE) key until the symbol
“
nF
”
is displayed.
3. Insert the red cable into input terminal VHz%CAPTemp
and the black cable
into input terminal COM.
4. Press
the
REL/
key before carrying out measurements.
5. Position the leads at the ends of the capacitor to be tested, respecting, if necessary,
the positive (red cable) and negative (black cable) polarity (see Fig. 27). The display
shows the value of capacitance. The bargraph is not active in this function.
6. The
message
“
O.L.
”
indicates that the value of capacitance exceeds the maximum
measurable value.
7. To use the HOLD and REL function, see § 4.2.
8. To use the internal functions, see § 4.3.