Assembly Procedure
Document Ref:
AP3607157
Issue:
01
Issue Date:
19 December 2019
Page 7 of 21
Drawing:
Shown on Works Order
Description:
A2A SPACER AUSTRALIA
SI No: N/A, Issue 2, 26 Jun 2017
50
ASSY
Valve tests
Use Piston Pulmonary Waveform Generator.
Insert mouthpiece assembly with two way valve in to pulmonary waveform generator as shown below
Set the operating program as shown below:
F-T 26 waveform
Waveform number 11
Inverse-continue flow type
When you click the fire piston button, this window will appear on the screen.
Set the number of cycles to 2
While the waveform generator is firing, look at the valve to ensure it is opening, as below.