Adapter usage (option)
1. When optional AC adapter should comply with the requirement of
IEC 60601-1:2005. Furthermore all configurations shall comply
with the requirements for medical electrical systems (see IEC
60601-1-1or clause 16 of the 3Ed. of IEC 60601-1, respectively).
Anybody connecting additional equipment to medical electrical
equipment configures a medical system and is therefore
responsible that the system complies with the requirements for
medical electrical systems. Attention is drawn to the fact that local
laws take priority over the above mentioned requirements. If in
doubt, consult your local representative or the technical service
department.
2. When using AC power, to avoid possible damage to the monitor,
use only the exclusive AC adapter that can be purchased from
authorized dealers. Other adapters may vary in output voltage and
polarities.
3. Insert the adapter plug into the hole on the backside of the unit
as picture.
4. Insert the other side of the adapter into the outlet with 100-240V.
5. To remove the AC adapter, disconnect the adapter plug from the
outlet first and then disconnect the cord from the unit’s socket.
Adapter technical features:
Output voltage: 6V
5%
Max. output current: At least 600 mA
Output plug polarity: <+> inner
External diameter: 5.5mm 0.1mm
Internal diameter: 2.1mm 0.1mm