Connecting Channels to Common for Making
Measurements
•
Closing any channel on the multiplexer module connects the
channel’s High (H) terminal to the Channels H Common Terminal
and the H terminal of the Analog Bus connector.
•
Closing the DT Tree Switch connects any channel on the multiplexer
module to the DT Tree Switch terminals.
•
Closing the DT Tree Switch connects any channel on the multiplexer
module to the I+, I- , and IG terminals of the Analog Bus connector.
•
All channels are referenced to one Low (L) and one Guard (G)
Common that are normally connected to each other. To isolate the
Low and Guard Commons, remove jumper J01.
Example: Connect
Channel 02 to
Common for
Voltage
Measurement
Figure 3-1 shows how to connect channel 02 of a single module switchbox
to the Common Terminals for a voltage measurement. Execute:
CLOS (@102)
102 closes channel 02; the first
number (1) is the card number
Figure 3-1. Connecting Channel 02 for a Voltage Measurement
26 Using the HP E1346A Multiplexer Module
Chapter 3