4-17
Reference
Cleaning Connections for Accurate Measurements
C A U T I O N
Hewlett-Packard strongly recommends that index matching compounds
not
be
applied to their instruments and accessories. Some compounds, such as gels,
may be difficult to remove and can contain damaging particulates. If you think
the use of such compounds is necessary, refer to the compound manufacturer
for information on application and cleaning procedures.
To clean a non-lensed connector
C A U T I O N
Do not use any type of foam swab to clean optical fiber ends. Foam swabs can
leave filmy deposits on fiber ends that can degrade performance.
1
Apply pure isopropyl alcohol to a clean lint-free cotton swab or lens paper.
Cotton swabs can be used as long as no cotton fibers remain on the fiber end
after cleaning.
Table 4-2. Cleaning Accessories
Item
HP Part Number
Isopropyl alcohol
8500-5344
Cotton swabs
8520-0023
Small foam swabs
9300-1223
Compressed dust remover (non-residue)
8500-5262
Table 4-3. Dust Caps Provided with Lightwave Instruments
Item
HP Part Number
Laser shutter cap
08145-64521
FC/PC dust cap
08154-44102
Biconic dust cap
08154-44105
DIN dust cap
5040-9364
HMS10/HP dust cap
5040-9361
ST dust cap
5040-9366
Содержание HP 83491A
Страница 1: ...User s Guide HP 83491 2 3A Clock Recovery Modules ...
Страница 4: ...iv General Safety Considerations ...
Страница 6: ...Contents 2 Contents ...
Страница 7: ...1 To install the module 1 3 To connect cables to an HP 83492A 1 7 Installation ...
Страница 14: ...1 8 Installation Installation ...
Страница 34: ...3 8 Specifications and Regulatory Information Declaration of Conformity Declaration of Conformity ...
Страница 60: ...4 26 Reference Hewlett Packard Sales and Service Offices ...