4-11
Reference
Cleaning Connections for Accurate Measurements
tions take repeatability uncertainty into account?
• Will a connector degrade the return loss too much, or will a fusion splice be re-
quired? For example, many DFB lasers cannot operate with reflections from
connectors. Often as much as 90 dB isolation is needed.
Figure 4-3. Basic components of a connector.
Over the last few years, the FC/PC style connector has emerged as the most
popular connector for fiber-optic applications. While not the highest perform-
ing connector, it represents a good compromise between performance, reli-
ability, and cost. If properly maintained and cleaned, this connector can
withstand many repeated connections.
However, many instrument specifications require tighter tolerances than most
connectors, including the FC/PC style, can deliver. These instruments cannot
tolerate connectors with the large non-concentricities of the fiber common
with ceramic style ferrules. When tighter alignment is required, HP instru-
ments typically use a connector such as the Diamond HMS-10, which has con-
centric tolerances within a few tenths of a micron. HP then uses a special
universal adapter, which allows other cable types to mate with this precision
connector. See
.
Содержание HP 83491A
Страница 1: ...User s Guide HP 83491 2 3A Clock Recovery Modules ...
Страница 4: ...iv General Safety Considerations ...
Страница 6: ...Contents 2 Contents ...
Страница 7: ...1 To install the module 1 3 To connect cables to an HP 83492A 1 7 Installation ...
Страница 14: ...1 8 Installation Installation ...
Страница 34: ...3 8 Specifications and Regulatory Information Declaration of Conformity Declaration of Conformity ...
Страница 60: ...4 26 Reference Hewlett Packard Sales and Service Offices ...