E
How
to
clean
instruments
with
a
recessed
lens
interface
Light
dirt
If
you
nd
any
particles
of
dirt
on
the
xed
connector
interface
,
or
on
the
input
of
the
optical
block,
clean
them
with
compressed
air
.
This
cleaning
method
is
gentle
to
your
optical
device
and
is
an
ideal
approach
if
there
is
no
fat
on
the
surface
of
your
device
.
Heavy
dirt
Using
compressed
air
,
blow
away
larger
particles
of
dirt.
Clean
the
recessed
lens
interface
with
a
dry
cotton
swab
by
rubbing
gently
over
its
surface
.
Then
blow
away
any
remaining
laments
left
by
the
swab
.
Isopropyl
alcohol
should
only
be
used
if
there
is
fat
on
the
lens
.
In
this
case
,
choose
a
new
swab
and
moisten
it
with
alcohol.
Remove
the
dirt
by
gently
rotating
the
swab
using
light
pressure
.
Then
select
a
fresh,
dry
swab
to
wipe
away
the
alcohol.
Blow
away
any
remaining
lament
with
compressed
air
.
Caution
Do
not
press
the
swabs
too
hard
onto
the
optical
surface
of
your
device
,
as
this
may
produce
scratches
and
your
recessed
lens
could
be
misaligned.
Alcohol
will
dissolve
some
types
of
dirt.
If
this
then
forms
a
deposit
on
the
edge
of
the
optical
input
it
cannot
be
removed.
Y
ou
should
therefore
only
use
alcohol
if
absolutely
necessary
.
Other
solvents
,
like
A
cetone
,
should
also
not
be
used,
as
they
may
damage
your
optical
device
and
reduce
its
performance
.
Cleaning
Procedures
E-7