D
I.
A
ccuracy
T
est
1.
On
both
power
meters:
a.
Set
the
calibration
factor
to
zero
.
b.
Set
the
wavelength
to
the
wavelength
of
the
laser
source
.
c.
Set
the
instruments
into
MEAS
mode
.
Make
sure
A
UTOranging
is
o
and
select
the
-20dBm
range
.
d.
Set
the
averaging
time
to
500ms
.
2.
Make
sure
that
the
HP
8156A
output
is
disabled.
ZERO
the
DUT
.
3.
Connect
the
HP
8156A
output
cable
to
the
standard
head
and
ZERO
the
standard
meter
.
4.
Enable
the
HP
8156A
output.
Change
the
HP
8156A
attenuation
until
the
power
meter
standard
displays
10.00W
.
5.
Connect
the
attenuator
output
cable
to
the
DUT
.
Check
the
DUT
display
and
note
the
result
in
the
test
record.
6.
Repeat
steps
1
to
5
at
the
second
wavelength
with
the
corresponding
source
.
II.
Linearity
T
est
1.
Make
sure
that
the
HP
8156A
output
is
disabled.
ZERO
the
DUT
.
2.
Enable
the
HP
8156A
output.
3.
On
the
DUT
,
switch
o
autoranging
and
select
the
-20dBm
range
.
Set
the
display
to
show
results
in
dBm.
4.
Alter
the
attenuation
until
the
DUT
displays
-17.4dBm.
5.
Press
4
Disp! Ref
5 ,
then
4
dB
5 .
6.
Press
4
Up
5
to
select
the
-10dBm
range
.
7.
Note
the
deviation,
displayed
in
dB
,
as
R1
on
the
test
record.
a.
In
the
-10dBm
range
.
Set
the
display
to
show
results
in
dBm.
b.
Alter
the
attenuation
until
the
DUT
displays
-7.4dBm.
c.
Press
4
Disp! Ref
5 ,
then
4
dB
5 .
d.
Select
the
0dBm
range
.
e.
Note
the
deviation,
displayed
in
dB
,
as
R2
on
the
test
record.
HP
81536A
P
erformance
T
ests
D-5