To test the auxiliary power
The +5 V auxiliary power is protected by a current overload protection
device. If the current on pins 1 and 39 exceeds 0.33 amps, the circuit will
open. When the short is removed, the circuit will reset in approximately 1
minute. There should be +5 V after the 1 minute reset time.
Equipment Required
Equipment
Critical Specifications
Recommended
Model/Part
Digital Multimeter
0.1 mV resolution, better
than 0.005% accuracy
HP 3478A
•
Using the multimeter, verify the +5 V on pins 1 and 39 of the probe
cables.
Troubleshooting
To test the auxiliary power
13-17
Содержание HP 1660CP Series
Страница 4: ...iv ...
Страница 16: ...xvi ...
Страница 17: ...1 Logic Analyzer Overview ...
Страница 26: ...1 10 ...
Страница 27: ...2 Connecting Peripherals ...
Страница 36: ...2 10 ...
Страница 37: ...3 Using the Logic Analyzer ...
Страница 58: ...3 22 ...
Страница 59: ...4 Using the Trigger Menu ...
Страница 76: ...4 18 ...
Страница 77: ...5 Using the Pattern Generator ...
Страница 126: ...5 50 ...
Страница 127: ...6 Triggering Examples ...
Страница 158: ...6 32 ...
Страница 159: ...7 File Management ...
Страница 163: ...Saving the System Configuration for Programmatic Control File Management To save a configuration 7 5 ...
Страница 168: ...7 10 ...
Страница 169: ...8 Reference ...
Страница 239: ...9 System Performance Analysis SPA Software ...
Страница 258: ...SPA Time Interval System Performance Analysis SPA Software SPA measurement processes 9 20 ...
Страница 269: ...10 Concepts ...
Страница 299: ...11 Troubleshooting ...
Страница 313: ...12 Specifications ...
Страница 322: ...12 10 ...
Страница 323: ...13 Operator s Service ...
Страница 331: ...Troubleshooting Flowchart 2 Troubleshooting To use the flowcharts 13 9 ...
Страница 340: ...13 18 ...
Страница 344: ...Glossary 4 ...