Installation Note E4400-90252
3
Standard ESG-D “B” to Option UN9 Upgrade Kit
Introduction
The combination of Option UN8 (Real-Time I/Q Baseband Generator with TDMA Standards) and
Option UN9, which provide an additional 7 M of data pattern RAM is added to the standard ESG-D
Model B Series Signal Generator.
To install this kit, the following major steps need to be performed.
1. Install Option UN8 and UN9 hardware.
2. Change cable assemblies W25 and W26.
3. Install the new firmware.
4. Verify the firmware.
5. Recalibrate and reassemble the RF signal generator.
After Option UN9 (UN8 + UN9) has been installed, the following major steps need to be performed to
verify the operation of the signal generator.
1. Activate Option UN9 (UN8 + UN9).
2. Internally, recalibrate the instrument.
Product Affected: . . . . . . . . . . . . . . . . . . . . . . . . .
ESG-D Series Signal Generators
Models Affected: . . . . . . . . . . . . . . . . . . . . . . . . . . .
HP E4430B, HP E4431B, HP E4432B, and
HP E4433B
Serial Numbers: . . . . . . . . . . . . . . . . . . . . . . . . . . .
All
Options: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
( )
Compatibilities: . . . . . . . . . . . . . . . . . . . . . . . . . . .
( )
To Be Performed By: . . . . . . . . . . . . . . . . . . . . . .
(X) HP Service Center
(X) Personnel Qualified by HP
(X) Customer
Estimated Installation Time: . . . . . . . . . . . . . . . .
2.0 hours
Estimated Verification Time: . . . . . . . . . . . . . . . . .
0.1 hour
Kit Part Number
Description
E4400-60185
Install Option UN9 (UN8 + UN9).