5-2
HP 8970B Option 020 Service Manual Supplement
Fault-Finding Tips
General
General
The Option 20 front end is relatively easy to fault find because it is a combination of
defined RF components.
Instead of using the noise source it is possible to use a signal generator to inject a signal
and then trace the signal levels through the RF paths to the 300 MHz second IF using a
spectrum analyzer. It is much easier to do it this way because the signals can be seen on
the spectrum analyzer display.
Filter (0955-0634)
The specifications on this part are:
The best way to measure this part is to use a network analyzer. An alternate way is to use
a spectrum analyzer, e.g. 8593E with a tracking generator or a separate sweeper.
Verify over the frequency ranges above that the insertion losses are correct.
Table 5-1: 2050 MHz Low Pass Filter Specifications
Frequency
Insertion Loss
10 MHz to 2050 MHz
< 0.5 dB
2.35 GHz to 3.85GHz
> 35 dB
3.85 GHz to 13.0 GHz
> 65 dB
13.0 GHz to 26.5 GHz
> 50 dB
Содержание 8970B
Страница 1: ...HP 8970B Option 020 Service Manual Supplement ...
Страница 2: ......
Страница 3: ...Service Manual Supplement HP 8970B Option 020 HP Part no 08970 90115 Edition 1 May 1998 ...
Страница 8: ...Contents 2 HP 8970B Option 020 Service Manual Supplement ...
Страница 9: ...1 Block Diagram and Parts List ...
Страница 14: ...1 6 HP 8970B Option 020 Service Manual Supplement Block Diagram and Parts List Material List ...
Страница 15: ...2 Theory of Operation ...
Страница 17: ...3 Adjustments ...
Страница 20: ...3 4 HP 8970B Option 020 Service Manual Supplement Adjustments 600 MHz SAW Oscillator Frequency Adjustment ...
Страница 21: ...4 Board Differences ...
Страница 25: ...5 Fault Finding Tips ...
Страница 31: ...6 Board Details ...
Страница 33: ...HP 8970B Option 020 Service Manual Supplement 6 3 Board Details Figure 6 2 600 MHz Schematic ...
Страница 39: ......