4-14
Programming
Common Commands
*ESR (Event Status Register)
The *ESR query returns the value of the Standard Event Status Register.
When you read the Event Status Register, the value returned is the total of the
bit weights of all of the bits that are set to one at the time you read the byte.
Table 4-2 shows each bit in the Event Status Register and its bit weight.
Reading the register clears the Event Status Register.
Usage:
HP-IB only
Query Syntax:
*ESR?
Returned Format:
<status><NL>
Where:
<status> ::= 0 to 255 (integer–NR1 format)
Example:
OUTPUT 711;"*ESR?"
ENTER 711;Event
PRINT Event
1
2
NOT USED
0
1
OPC – Operation Complete
Table 4-1. Standard Event Status Enable Register
(High–Enables the ERS bit)
Bit
Bit Weight
Enables
Table 4-2. Standard Event Status Register
Bit
Bit Weight
Condition
7
128
PON – Power On
6
64
URQ – User Request
5
32
CME – Command Error
Содержание 86060C Series
Страница 1: ...User s Guide HP 86060C Series Lightwave Switches ...
Страница 18: ...1 12 General Information Specifications and Regulatory Information ...
Страница 46: ...3 6 Using the Switch Changing Switch Position Changing Switch Position ...
Страница 93: ...5 Spare Channel Replacement Procedure 5 4 Electrostatic Discharge Information 5 7 Servicing ...
Страница 98: ...5 6 Servicing Spare Channel Replacement Procedure Figure 5 2 Top view of the HP 86062C lightwave switch ...
Страница 104: ...Index 4 Index ...